Maintenance and Service
140
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Cleaning the Instrument Electrodes
The electrodes should be cleaned daily. Also clean the electrodes
after running any diagnostic test using Test Chip D.
To clean the electrodes, wipe down using the provided lint free
swab damped in DI water.
Cleaning the Microplate Carrier
The microplate carrier can be cleaned with DI water or isopropyl
alcohol as needed.
Cleaning the Touch Screen
Clean the touch screen as necessary.
Glass cleaner, isopropyl alcohol, or ethanol can be used on the
touch screen. When cleaning, spray cleaner onto a cloth and wipe
the screen with the cloth.
WARNING
Do not spray cleaner directly onto the touch screen. Always spray
cleaner onto a cloth and then wipe the screen with the cloth.
Cleaning the Optics Lens
Inspect the optics lens daily and clean as necessary.
Clean the optics lens gently with an optical wipe dampened with DI
Water.
WARNING
•
Do not spray cleaner directly into the instrument. Always spray
cleaner onto a cloth and then wipe with the cloth.
•
Clean the optical lens with an optical wipe and DI water only.
Using other cleaning materials, or cleaning vigorously, can
damage the lens.