Instrument Safety
11
P/N CLS151164 Rev. BLabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Electrical Safety
The LabChip GX Touch/GXII Touch is powered by a UL/CSA/VDE
approved 100-240 VAC, 50/60 Hz input, 5, 15, 24 VDC output
power supply. Additionally, the LabChip GX Touch/GXII Touch High
Voltage circuitry is current-limited to non-hazardous levels. Users
should observe the following:
WARNING
Do not open the instrument enclosure. There are no user
serviceable parts inside.
The wall outlet or the power cable connector on the back of the
instrument should be accessible after the system's installation, to
enable trained service personnel to safely disconnect power from
the system during servicing.
The computer supplied with the LabChip GX Touch/GXII Touch
instrument has internal lithium batteries. Batteries should not be
incinerated.
WARNING
Danger of explosion if battery is incorrectly replaced. Replace only
with the same or equivalent type recommended by the
manufacturer's instructions.
Power Cord Selection
United States and Canada
The LabChip GX Touch/GXII Touch instrument is shipped with a
NEMA 5-15 / IEC 320 power cord. If the power cord needs to be
replaced, substitute power cords must be UL Listed, Type SJT or
equivalent, minimum No. 18 AWG, 3-conductor with ground
conductor that for safety considerations should never be
disconnected or defeated. The cord’s plug to the wall must be a
three-pin grounding type connector with a NEMA 5-15P (15A,
125V) plug configuration. The cord’s connector at the unit must
conform to requirements for an EN 60 320/IEC 320 Standard Sheet
C13 connector.
The equipment is intended to be plugged into a standard NEMA 5-
15R receptacle in the wall.