Software Reference
109
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
EGram Tab
The EGram Tab is a visual representation of the data from each well
as an electropherogram. A single well or multiple wells can be
viewed in the EGram tab.
Use the EGram Tab to view data from completed wells. To change
the view in the EGram tab, see
Viewing Graphs in the Egram Tab
. To open the EGram tab, touch the
Views
button on the
and then touch the
EGram
button.
Figure 42. EGram Tab
Touch and drag over a region on the electropherogram to zoom in.
Touch the Unzoom button or the Unzoom All button to zoom out.
Graph data can be exported to a graphic file at the end of the run by
choosing
Auto Export
on the
for details.)
To show or change the labels on the peaks in the graph, show the
data points on the graph, show peak baselines, show smears, or
change the graph colors, see
.
While a run or calibration is in progress, the
displays
the raw (unanalyzed) data as it is being read from the chip.