Lambda Spectrometer Quick Setup Guide
Installing the Spectrometer
WARNING
As the Lambda 25, 35, 45 weighs approximately 26 kg, we recommend that the instrument is
lifted by 2 adults, and that it is lifted by the base of the instrument.
1.
Create a space on the bench in your laboratory.
Dimensions of the spectrometer:
Width 56 cm
(22.05 inches)
Length 65 cm
(25.59 inches)
Height 26 cm
(10.24 inches)
NOTE:
Remember to leave space for your PC on the right hand side of the spectrometer.
2.
Unpack and check all components:
•
Lambda Spectrometer
•
2 Single Cell Holders
•
Hexagonal Wrench
•
Screwdriver
•
RS232 Cable (09410022)
•
Line Power Cord
•
Spare Fuse
•
Software Kit
3.
Examine the components for any signs of damage during shipment. Make sure terminals, fuse holders, etc.
are not damaged. Open and close the sample compartment cover, checking that it moves freely. If the
instrument has been damaged, do not switch it on. Please contact your local PerkinElmer service
representative.
WARNING
Electrical Hazard:
To prevent potential injury to yourself and damage to the instrument, first
make the electrical connections between the instruments in the system before connecting to
the line power supply. The spectrometer automatically adjusts to the correct operating voltage.
Before starting the instrument for the first time, make sure that the correct fuse is fitted to your
line power supply. If you have a different fuse, change it for the correct one.