Lambda Spectrometer Quick Setup Guide 

Installing the Spectrometer 

 

 

 

WARNING 

As the Lambda 25, 35, 45 weighs approximately 26 kg, we recommend that the instrument is 

lifted by 2 adults, and that it is lifted by the base of the instrument. 
 

 

1.

 

Create a space on the bench in your laboratory. 

Dimensions of the spectrometer:   

Width  56 cm  

(22.05 inches) 

Length 65 cm  

(25.59 inches) 

Height  26 cm  

(10.24 inches) 

 

NOTE:

 Remember to leave space for your PC on the right hand side of the spectrometer.

 

 

2.

 

Unpack and check all components: 

 

Lambda Spectrometer 

 

2 Single Cell Holders 

 

Hexagonal Wrench 

 

Screwdriver 

 

RS232 Cable (09410022) 

 

Line Power Cord 

 

Spare Fuse 

 

Software Kit 

 

3.

 

Examine the components for any signs of damage during shipment. Make sure terminals, fuse holders, etc. 
are not damaged.  Open and close the sample compartment cover, checking that it moves freely. If the 

instrument has been damaged, do not switch it on. Please contact your local PerkinElmer service 
representative. 
 

 

 

WARNING 

Electrical Hazard:

 To prevent potential injury to yourself and damage to the instrument, first 

make the electrical connections between the instruments in the system before connecting to 

the line power supply. The spectrometer automatically adjusts to the correct operating voltage. 

Before starting the instrument for the first time, make sure that the correct fuse is fitted to your 

line power supply. If you have a different fuse, change it for the correct one. 

 