Clarus SQ 8 MS Hardware Guide
166
Spectral Related
Problem
Probable Cause
Solution
Noisy spectra.
Dirty source.
Remove and clean the source.
Peak detection threshold set too low.
Raise the thresholds.
Multiplier voltage set too high.
Lower the multiplier voltage.
Spectrum distortion.
Scanning too fast or slow.
Reset the scan rate.
Incorrect Isotope ratios. Poor calibration.
Recalibrate.
Incorrect tune.
Retune.
Defective filament.
Replace filament.
Air leak.
Find and fix.
Missing Isotopes in
spectrum.
Bad calibration.
Recalibrate.
Poor tuning.
Retune.
Dirty source.
Clean source.
Sample too weak.
Use a higher sample
concentration.
Peak detection thresholds set too
high.
Lower the thresholds.
Multiplier voltage set too low.
Raise the multiplier voltage.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......