332
Troubleshooting
Pump Problems
Table 6-18.
Pump Problems
Problem
Possible Cause(s)
Corrective Action
Pump rollers stick
A spill may have occurred or
the pump head may be worn
and may need to be replaced.
Clean the pump head, or replace if
necessary.
Liquid is not
flowing freely
The tension on the pump
tubing is too high or low.
Tubing may not be installed
correctly in the channel.
Tubing may be worn.
Use the adjustment screw to adjust the
tension. See Chapter 5, Maintenance
for more information.
Reinstall tubing.
Replace tubing.
Pump tubing is
stretched on one end
and slack on the
other end
The tension on the pump
tubing is too tight.
Rollers may be stuck.
Tubing may be worn.
Loosen the adjustment screw.
Check the rollers in the pump head to
see if they roll freely.
Replace tubing.
Pump makes
excessive noise
A mechanical problem may
have occurred.
Call a PerkinElmer Service
Representative.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......