118
Installation
Connecting the Gases and Cooling Water
After the instrument has been moved into its position, it can be connected to the
various services in the laboratory.
Connecting the Gases
Use two wrenches when making Swagelok connections. Use a space collar (Part
No. 09920584, supplied in the hose kit) at each Swagelok connection. Tighten the
Swagelok nut until you can no longer turn the knurled space collar.
Only if you
cannot locate a space collar,
tighten the nut until it is finger-tight, then turn it an
additional 1/4 turn with a wrench. Pre-swaged fittings should be turned 1/8 turn
past finger-tight. Do not over-tighten fittings. Check for leaks with all the
connections made.
Connect the gases to the instrument as described in the following procedures.
Connecting the Nitrogen Supply
1. Connect the instrument to the nitrogen supply using the nitrogen hose (Part No.
N0690275, uses 1/4-in. Swagelok fitting) using a 1/4-in. space collar (Part No.
09920584) on both ends of the tubing. Refer to the following figure.
2. Set the nitrogen supply regulator between 275 kPa and 825 kPa (2.75 to 8.25
bar or 40-120 psig).
3. Check for leaks using a proprietary leak testing fluid for high purity gas lines.
Note
All gas hose connections use 1/4 inch Swagelok fittings.
Caution
Attention
The pneumatic tubing kinks easily.
Install it so that it is less likely to twist, fold and
kink
.
Les kinks de tubes pneumatiques facilement. Installez-le de sorte qu'il est moins
susceptible de se tordre, plier et kink.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......