![PerkinElmer AVIO 200 Hardware Manual Download Page 254](http://html1.mh-extra.com/html/perkinelmer/avio-200/avio-200_hardware-manual_1530192254.webp)
Maintenance
252
Adjusting the Pump Tubing
Adjust the tension on the sample and drain tubing, one at a time, using deionized
water as follows:
Adjusting the Sample Tubing
1. Make sure the sample tubing is centered in the channel, then place the clamp
around the tubing and lower the clamp arm on onto the pump. Turn the
adjustment screws until locked in place.
2. Place the sample capillary (SAMPLE IN side) in a container of deionized water.
3. Set the pump speed to 1.0 mL/min and start the pump.
4. Dip the tubing in DI water a few times to let air gaps form. As you watch the air
gaps form loosen the adjustment screw in a counterclockwise direction until the
air gaps stop or do not flow smoothly.
5. Tighten the adjustment screw
slowly
(turn clockwise) until droplets just start
forming at the end of the SAMPLE OUT side of the sample tubing.
6. Tighten the adjustment screw until you see a smooth flow of bubbles (3 turns on
the peristaltic pump). Add an additional three more turns.
7. Turn off the pump. The tension is now correctly set. No bubbles should be seen
in the sample tubing once the sample tube is kept in the solution.
Adjusting the Drain Tubing
1. Make sure the drain tubing is centered in the channel without disturbing the
sample tubing, then place the clamp around the drain tubing and lower the clamp
arm.
2. Loosen the adjustment screw (turn counterclockwise) for the drain tubing
channel so that the lever can be easily opened and closed with little resistance.
3. Set the pump speed to 1.0 mL/min and start the pump.
4. Dip the tubing in DI water a few times to let bubbles form. As you watch the
bubbles form loosen the adjustment screw in a counterclockwise direction until
the bubbles stop or do not flow smoothly.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......