Maintenance
251
Installing the Drain Tubing
Install the drain tubing on an open channel of the peristaltic pump as follows:
1. Pre-stretch a piece of 1.14 mm (0.045 in.) pump drain tubing.
2. Carefully stretch the new tubing around the pump head and place the clips on the
pump tubing into the slots, using the slots in the same position on each side of
the pump head. Be careful not to twist or kink the pump tubing during this
process.
3. Attach the Teflon tubing to the pump tubing for the drain.
Make sure you install the tubing on the pump so that the waste liquid is
pumped
out
of the spray chamber. See Figure 5-36.
If the tubing is installed
in the wrong direction, the spray chamber will flood.
4. For the tubing that leads to the drain bottle, use one piece of 1.07 mm i.d. (0.042
in.) Teflon tubing (Part No. 09985708). Bevel one end by cutting it at an angle.
5. Insert the Teflon tubing into the other side of the pump tubing. Lead this tubing
to the drain bottle. Refer to Figure 5-36.
6. Adjust the tubing as described in the next procedure.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......