![PerkinElmer AVIO 200 Hardware Manual Download Page 246](http://html1.mh-extra.com/html/perkinelmer/avio-200/avio-200_hardware-manual_1530192246.webp)
Maintenance
244
Cleaning the Nebulizer
For the longest life and best performance, wash your nebulizer by simply running
water as a sample for 10 minutes at the end of the day before shutting down the
plasma. Any other form of washing is usually unnecessary.
DO NOT
wash your
nebulizer in acid or solvents to "prevent salt build up".
DO NOT
use surfactants as
they will destroy the surface tension at the tip and performance.
Peristaltic Pump
The peristaltic pump requires routine maintenance to ensure that the system will run
properly. The pump tubing should be replaced when flat spots develop (on a daily
basis or after approximately eight hours of use). When replacing the pump tubing,
the drain tubing should be replaced at the same time since uneven wear can cause the
spray chamber to back up.
Replacement Pump Tubing
Replace the pump tubing when flat spots develop (after approximately eight hours
of use). When replacing the pump tubing, both the sample and drain tubing should
be replaced at the same time since uneven wear can cause the spray chamber to back
up.The following table gives part numbers for pump tubing for the sample and drain.
Caution
Attention
The gas orifice is at the very tip of the nebulizer. It is made of Teflon which
is
VERY SOFT
. This tip is very easily damaged and should
NEVER
be
touched with fingers, tissues, or anything else. If the tip is accidentally
touched and the nebulizer continues to operate, then it is still functional,
and its use can be safely continued.
L'orifice de gaz est à la pointe du nébuliseur. Il est fait de téflon qui est très
doux. Cette astuce est très facilement endommagé et ne doit
JAMAIS
être
touché avec les doigts, les tissus, ou toute autre chose. Si la pointe est
accidentellement touché et le nébuliseur continue à fonctionner, alors il est
encore fonctionnel, et son utilisation peut se poursuivre en toute sécurité.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......