Maintenance
235
Installing the GemCone Nebulizer on the Cyclonic Spray Chamber
1. Before inserting the GemCone nebulizer into the chamber, connect the argon gas
fitting to the nebulizer threads. First, tighten the gas fitting by hand. Tighten the
fitting another 1/8 of a turn using a 9/16-in. open-end wrench. Support the
nebulizer with a back-up wrench when tightening or loosening fittings. Do
not
overtighten the fitting as this may cause leaking and poor ICP performance.
2. Connect the capillary tubing to the nebulizer. Be sure not to crimp or bend the
tubing as you insert it. Crimps may reduce flow to the tip of the nebulizer or may
cause air to enter the liquid stream during operation.
3. Insert the nebulizer into the cyclonic spray chamber as shown in the next figure.
The nebulizer may be inserted to its fullest extent.
Figure 5-28
Connecting the GemCone nebulizer to the cyclonic spray chamber.
Item
Description
Item
Description
1
Cyclonic Spray Chamber
4
GemCone Nebulizer Assembly
2
Tubing Assembly for the
GemCone Nebulizer (Part No.
N0770336)
5
Teflon Tubing-1 inch (Part No.
09985708)
3
Argon Gas Fitting
6
Capillary Tubing (Part No. 09908265)
1
2
3
5
6
4
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......