Maintenance
218
GemTip Cross-Flow Nebulizer
The following procedures cover the maintenance of the GemTip Cross-Flow
nebulizer supplied with the instrument.
To check the performance of the nebulizer, aspirate a standard and note the intensity
reading. If the intensity is significantly lower than it should be, first check that:
•
The pump tubing is in good condition (has no flat spots).
•
The sample capillary is in good condition (has no kinks) and is clean.
•
The plasma viewing position is set properly (using the software's alignment
function).
Also try flushing out the nebulizer by aspirating for several minutes using either:
•
a 1% acid solution, if you have been analyzing aqueous solutions, or
•
a solvent, if you have been analyzing organic solutions.
Recheck the intensity of the standard. If the intensity is still too low,
you should
check the nebulizer spray pattern. Before removing the nebulizer/end cap to check
the spray pattern, be sure to pump deionized water for several minutes to thoroughly
flush it out.
When checking the spray pattern of the nebulizer pump only use
deionized water. Do
not
use any other solution.
Avertissement:
Lors de la vérification du motif de pulvérisation de la
pompe nébuliseur utiliser uniquement de l'eau déminéralisée.
Ne pas
utiliser toute autre solution.
!
Warning
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......