3. See the following procedure on installing a new shear gas nozzle or replacing the
existing shear gas nozzle.
Returning the Shear Gas Nozzle
Use this procedure to install a new shear gas nozzle or returning the existing shear
gas nozzle if you had to remove it in order to replace either the axial or radial purge
window.
1. Place the shear gas assembly back into position. Insert the two screws into the
slots on the bracket. Make sure the screws are located at the top of the slot before
tightening with the T20 Torx screwdriver (
2. Reinstall the torch assembly; see the section
earlier in this chapter.
3. Align the torch and perform an axial and radial alignment; refer to the procedure
Setting the Torch Viewing Position
Note
Before you perform an analysis at wavelengths below 190 nm, purge the
optics overnight on high purge. For other analyses, purge the system until
you obtain a steady signal.
Table 5-4.
Replacement Parts: Purge Components
Part No.
Description
09921036
O-Ring (for Dual View Radial Purge Tube)
09921057
Purge Outlet O-Ring
09992731
Axial Purge Window
N0690672
Short Radial Purge Window
B0810377
Long Radial Purge Window
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......