Maintenance
193
manufacturer's safety recommendations. These recommendations are normally
provided in a material safety data sheet (MSDS) supplied with the chemical.
If you have been analyzing organic solutions, the spray chamber can be cleaned
using a solvent or diluted soap solution. Be sure to inspect the torch and injector daily
for carbon buildup.
O-rings may be cleaned with soap and water. An ultrasonic bath may be used.
Replace if cracked or worn. A torch O-ring kit is available. See the list of
replacement parts at the end of this chapter.
Replacing the Scott-Type Spray Chamber
See the procedure
Connecting the Scott Spray Chamber
Cyclonic Spray Chamber
The cyclonic spray chamber is a glass spray chamber. The cyclonic spray chamber
may be used with the concentric glass, low-flow GemCone, and high flow GemCone
nebulizers.
There are two cyclonic spray chambers. The baffled cyclonic spray chamber (Part
No. N0791352) is recommended for all aqueous and most organic solutions. For
really volatile organics, cyclonic spray chamber (Part No. N0777497) can be used.
Removing the Cyclonic Spray Chamber
1. Disconnect the tubing to the cyclonic spray chamber.
Handle the cyclonic spray chamber carefully to avoid breaking the
glassware and possible injury.
Avertissement: Manipuler avec soin la chambre de nébulisation
cyclonique pour éviter de casser la verrerie et des blessures.
!
Warning
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......