Maintenance
172
GemTip or GemCone
Nebulizer/End Cap
Assembly
Clean with soap and water; rinse thoroughly.
Concentric Glass
(MEINHARD)
Nebulizer
Do
not
use an ultrasonic cleaner. Do
not
use a cleaning
wire. See
Concentric Glass Nebulizer
later in this
chapter.
Mira Mist Nebulizer
For the longest life and best performance, wash your
nebulizer by simply running water as a sample for 10
minutes at the end of the day before shutting down the
plasma. Any other form of washing is usually
unnecessary. Please
DO NOT
wash your nebulizer in
acid or solvents to 'prevent salt build up'.
DO NOT
use
surfactants as they will destroy the surface tension at
the tip and performance.
The tip is very easily
damaged and should NEVER be touched with
fingers, tissues, or anything else
.
Spray Chamber
By handling the spray chamber oil and dirt can
accumulate on it and this will affect performance. To
assure proper wetting wipe the spray chamber with
acetone to remove all oils and dirt.
Pump Head
Remove and clean with water or a mild solvent.
Purge Viewing
Windows/Tubes
Clean with deionized water and dry with a soft cloth.
To clean difficult stains, refer to the information in this
chapter.
Table 5-1.
Recommended Cleaning Procedure
(Continued)
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......