Maintenance
170
Nebulizer
Check the nebulizer spray pattern with deionized water. Clean or replace the
nebulizer as necessary.
Spray Chamber
Check for leaks around the nebulizer and drain fitting. Inspect the spray chamber for
deposits and check the condition of the O-rings.
Peristaltic Pump
Check that the pump rollers are clean, not scored, free from deposits due to spills,
and move freely. Remove and clean the pump head if necessary. Replace the pump
rollers if necessary.
Periodically observe the drain tubing. Liquid should drain smoothly from the spray
chamber. If not, adjust the pump tension on the pump tubing.
Drains
Check the spray chamber drain fittings, tubing and connections and replace if
necessary.
General System Maintenance
•
Air filters:
clean or replace the spectrometer and generator air filters as
necessary.
•
Chiller:
Flush out the chiller every 6 months. Fill the chiller reservoir with the
coolant fluid that includes inhibitor (Part No. N0776200). The chiller reservoir
has a capacity of 4.2 liters (1.1 US gallons).
Refer to Chapter 4, Installation for more information on filling the chiller.
Caution
Attention
Do not use deionized or tap water in the chiller. Only use coolant fluid
that includes inhibitor (Part No. N0776200).
Ne pas utiliser déminéralisée ou l'eau du robinet dans le refroidisseur.
Utiliser uniquement du fluide de refroidissement qui comprend un
inhibiteur (N0776200).
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......