Installation
98
Setting Up the Computer and Printer
To install the ICP software, computer, and printer, refer to
Syngistix Software
Installation Guide
on Syngistix Document Pack CD.
Moving the Instrument
If you must move the instrument any great distance, especially if the instrument may
be subject to vibration or jolts,
contact a PerkinElmer service engineer to assist
you in moving your system
.
If you are only moving the instrument within the lab or to a laboratory nearby, you
may use the following procedure.
•
Remove the Quick-Change Torch module.
•
Remove the chiller.
•
Remove any accessories, for example an autosampler.
•
Disconnect all gas and water lines.
•
Remove the computer and printer.
•
Attach the handles.
Caution
Attention
I
f moving the Avio 200 will subject the instrument to any freezing temperatures you
must
contact a PerkinElmer service engineer to assist you in the move. The
PerkinElmer service engineer will flush all traces of cooling water from the RF
generator to prevent freezing of RF generator components.
Si le déplacement du Avio 200 assujettira l'instrument à des températures de
congélation, vous devez contacter un technicien de maintenance PerkinElmer pour
vous aider dans le mouvement. L'ingénieur de service PerkinElmer rincera toutes
les traces de l'eau de refroidissement du générateur RF pour empêcher le gel des
composants RF du générateur.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......