Laboratory Requirements
0993-5255
6
-15
Set the outlet gauge pressure for each gas to a value just below the maximum
value (
P
max
).
Never
set the outlet pressure to a value exceeding the maximum pressure
(
P
max
).
Table 6-2. Specifications and Requirements of the Furnace Gases
Gas
Purity
Outlet Gauge Pressure
Max.
Flowrate
kPa
bar
psig
L/min
P
min
P
max
P
min
P
max
P
min
P
max
1.2
Argon
99.996%
350
400
3.5
4.0
50.0
58.0
1.2
Nitrogen
99.996%
350
400
3.5
4.0
50.0
58.0
1.2
Special
particle free;
water free
350
400
3.5
4.0
50.0
58.0
1.2