Getting Started with the Detector
Instrument Specifications
1 - 20
Instrument Specifications
Table 1-2.
Specifications.
Specification
Description
Technical Specifications
Wavelength range
190-700 nm
Bandwidth
5 nm
Wavelength accuracy
+/- 1 nm
Optics
Dual beam
Sensitivity Range
0.0005 to 3.000 AUFS in 0.0001 increments from 0.0005 to 0.1,
and 0.01 increments above 0.1 AUFS.
Recorder output
10 mV full scale for any selected range.
Computer output
1 V/AU, 0.5 V/AU, 0.2 V/AU (software control)
Noise
< +/- 1 x 10
-5
AU, 210-280 nm, 1 sec. rise time, test flowcell.
Drift
< 1 x 10
-4
AU/hour, after warm-up
Light sources
Deuterium (190-360nm) or Tungsten (360-700nm) prefocused-no
adjustment required on replacement.
Leak detection
Heated thermistor sensor in glass envelope located in flowcell
drip pan
Display
Graphic LCD module 240 X 64 pixel. Series 200 standard.
Noise filter
Digitally controlled active filter with response (RSP) or peak
width (PW) selectable from 0.02 to 5.0 seconds in 8 steps.
Number of methods
20 in battery backed memory
Steps per method
20 maximum; multiple simultaneous events may occur in any
given step.
Method time
Up to 999.99 minutes per method.
Programmable functions from
the front keypad or Turbochrom
Time, wavelength, response time, auto-zero, relay closures,
method number, number of injections, (method execution via
sequence)
Scan
Stopped flow scan, scan speed selectable from 0.2 to 1.0 nm/sec.
External communications
RS-232 for control, down load of methods via Turbochrom and
upload of chromatographic data. Max. chromatographic data
rate is 50 data points per second.
Diagnostics
Self testing as per software diagnostic standards.
Microprocessor
Motorola 68332
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...