68153 01
www.perkinelmer.com
29
XRpad2 4336
D
IGIT
AL
I
MAGING
9.0
Inspection and Maintenance
It is important that the X-ray detector is used safely and as intended. Inspect the detector and its
accessories before use. If any problem is found during the inspection, correct the problem, and
take the measures indicated in this section. The X-ray detector and the Interface and Power Unit
do not have any internal exchangeable parts. Only the accessories listed in
can be
changed after consulting your with establishment’s safety representative. If the problem cannot
be corrected, contact your dealer, distributor, device manufacturer, or any PerkinElmer subsid-
iaries (regional service headquarters) listed on the last page of this document.
We recommend that records of the inspection be kept close to the detector. You can use copies of
the checklist in this section, or you can make your own copies of the checklist.
9.1
Daily Inspection
Perform the following inspection daily. If there is any problem, immediately ask your establish-
ment’s safety representative to contact your dealer, distributor, or device manufacturer.
9.1.1
Before Turning On the Power
WARNING
The X-ray detector must be repaired by PerkinElmer authorized
personnel only. Ignoring this warning may result in explosion, fire,
electric shock, or unknown hazards, which may result in severe
personal injury, death, or substantial product damage.
Caution
Inspect the X-ray detector before use. In addition, carry out prescribed,
regular inspections per the instructions in this manual.
Inspection
Result
Remedy
Date
/
Date
/
Date
/
Ca
bl
es
Check all cables, if
applicable, (Power and
communication tethered
cord, DC-cable, Ethernet
cable, Sync cable) to ensure
that they are not damaged
and the insulation is not
damaged.
Good/Bad Good/Bad Good/Bad Contact your dealer,
distributor, or device
manufacturer if there is a
problem.
Check all connector plugs
and locks to ensure they are
not loose.
Good/Bad Good/Bad Good/Bad Fully insert the cables and
lock them.