78 . RamanStation 400 Series Getting Started Guide
Fiber Optic Probe SOPs
NOTE: The SOPs included on the following pages are only appropriate for a standard fiber
optic probe. For details of how to acquire data using a triggered fiber optic probe, see
the
Raman Triggered Fiber Optic Probe
leaflet (L1321940).
SOP-6: Selecting the Fiber Optic Probe
1.
Ensure the RamanStation 400F is switched on and the fiber optic probe connected to
the instrument.
If not, follow SOP-1.
2.
Select
Raman Instrument
from the Setup menu in Spectrum.
The Setup Instrument tabs are displayed in the Dialog pane.
3.
Select the Setup Instrument Basic tab.
4.
Select the Fiber Optic Probe you want to use from the Accessory drop-down list.
The following warning message is displayed.
5.
Ensure that laser safety precautions are being observed, and then click Continue.
If you have any doubts about the required safety precautions click Cancel.
Laser light will be emitted from the probe during data collection and when the software is in
Monitor or Preview modes. When the laser is on and the shutter is open the Laser Status on
the Measurement bar will be updated to Class 3B,
. The probe can now be positioned
so as to achieve the best spectrum. See SOP-7.
Summary of Contents for RamanStation 400 Series
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 RamanStation 400 Series Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 RamanStation 400 Series Getting Started Guide ...
Page 35: ...Preparing Your Facility ...
Page 39: ...Installing Your RamanStation 400 Series ...
Page 52: ...52 RamanStation 400 Series Getting Started Guide ...
Page 53: ...Introduction to the RamanStation 400 Series ...
Page 59: ...Fiber Optic Probe ...
Page 66: ...66 RamanStation 400 Series Getting Started Guide ...
Page 67: ...Using the RamanStation 400 ...