![Perkin-Elmer RamanStation 400 Series Getting Started Manual Download Page 46](http://html1.mh-extra.com/html/perkin-elmer/ramanstation-400-series/ramanstation-400-series_getting-started-manual_1530158046.webp)
46 . RamanStation 400 Series Getting Started Guide
Attaching the Fiber Port Cover Assembly
NOTE: This section is only applicable if you have a RamanStation 400F.
The RamanStation 400F is classified as a Class 3B laser product according to the label
attached to the instrument. However if required, the RamanStation 400F can be classified as
a Class 1 laser product provided the following conditions are met:
•
No fiber optic probe is connected to the instrument
•
The fiber optic probe cover assembly (L1320263) is attached to the instrument and
completely encloses the EXC and COL couplers at the rear of the instrument. The EXC
and COL screw-in caps are fully screwed into their ports before the fiber optic probe
cover assembly is secured in place.
A Class 1 laser product is safe under all reasonably foreseeable conditions of use and may be
used in an open laboratory environment.
To attach the fiber port cover to the rear panel, fasten the supplied screw through the cover
into the hole near the COL coupler as shown in Figure 19.
Figure 19 Attaching the fiber port cover
Screw
Summary of Contents for RamanStation 400 Series
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 RamanStation 400 Series Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 RamanStation 400 Series Getting Started Guide ...
Page 35: ...Preparing Your Facility ...
Page 39: ...Installing Your RamanStation 400 Series ...
Page 52: ...52 RamanStation 400 Series Getting Started Guide ...
Page 53: ...Introduction to the RamanStation 400 Series ...
Page 59: ...Fiber Optic Probe ...
Page 66: ...66 RamanStation 400 Series Getting Started Guide ...
Page 67: ...Using the RamanStation 400 ...