Fault analysis
Analogue output
Possible cause
Removal
"
Bar display flashes
Response of the current output
depends on the setting in V0H7
V0H7=0 -10 % 2.4 mA or 3.8 mA
V0H7=1 110 % 22 mA
V0H7=2 HOLD last value is held
Error code in V9H0
E641 in V9H0
Echo too weak or foam
on the surface
• Which error code? see page 28
• Further action depends on the
error code
• Check sensor position
#
Measured value in
V0H0 too small
• Multiple echo? see
&
• Gas layering? Contact
Fuchs service line
• Check sensor position
Re-enter linearisation curve
see page 22 and page 23
Check values in V0H5 and V0H6
and re-enter if necessary
see page 24
$
Measured value in
V0H0 too large
Interference from internals in
measuring range?
Instrument mounted in nozzle?
• Check dimensions of nozzle
• Check sensor position
• Select application parameter
0 or 2 in V0H3 see page 21
• Carry out interference echo
suppression see page 25
yes
yes
20 mA
4 mA
t
D m/ft (V0H8)
expected
actual
Distance D in V0H8
too large?
Incorrect linearisation?
Incorrect current
output?
no
no
yes
yes
yes
20 mA
4 mA
t
D m/ft (V0H8)
expected
actual
Distance D in V0H8
too small?
Continued on page 30