27
e
14.5
Simulation
The simulation mode enables LUC-T functions
to be simulated and checked.
Errors and warnings in V9H0
• E613: Display during simulation. Returns to
normal operation after simulation.
Simulation off: V9H6: 0
• On power failure the instrument
automatically returns to normal operation!
Simulation of height
Simulation of current
Simulation of volume
14.6
Locking
Locking via the keyboard
When the instrument is locked via the
keyboard, both keyboard and display
parametration as well as all parametrations via
the handheld terminal etc. are blocked.
It can only be unlocked using the keyboard.
The matrix can be again locked once all
parameters have been entered.
• Locking by entering a three-figure code
number not equal 333.
• Locking by using the keyboard(see note on
locking via the keyboard).
#
VH
Entry
Text
1
V9H6
1
H
Simulation height
2
V9H7
e. g. 2.000
(6.600 ft)
H
Simulated height
(e. g. 2 m/6.6 ft)
3
V9H8
V0H0
Display of current (also shown on
bargraph)
Display of height, level or volume
4
V9H6
0
H
Simulation off
#
VH
Entry
Text
1
V9H6
3
H
Simulation current
2
V9H7
e. g. 14
H
Simulated current
(e. g. 14 mA)
3
V9H8
Display of current (also shown on
bargraph)
4
V9H6
0
H
Simulation off
#
VH
Entry
Text
1
V9H6
2
H
Simulation volume
2
V9H7
e. g.
100.0
(26.40
USgal)
H
Simulated volume
(e. g. 100 l /
26.4 USgal)
3
V9H8
V0H0
Display of current (also shown on
bargraph)
Display of volume (If no linearization
curve has been entered then volume
corresponds to level.)
4
V9H6
0
H
Simulation off
#
VH
Entry
Text
1
V9H9
e. g. 332
H
Locking
2
The number 332 is shown in V9H9.
All matrix fields are blocked except V9H9.
#
VH
Entry
Text
1
V9H9
333
H
Unlocking
2
The number 333 is shown in V9H9.
Matrix fields are no longer blocked.
V
V
H
H
+
+
V
V
H
H
P---
F---
0 2
0 2
P = Protect
Locking
display approx.
2 s
9999 displayed
in V9H9.
F = Free
Unlocking
display approx.
2 s
333 displayed in
V9H9