Prima SP2 Anaesthetic Machine Range
User Manual
Partnership for Life
دﯾﯾﺎﻣرﻓ ﮫﻌﺟارﻣ.WWW.ELM-FAN.IRتﯾﺎﺳ ﮫﺑ رﺗﺷﯾﺑ دراوﻣ دوﻠﻧاد یارﺑ
To download more items, go to ELM-FAN.IR
Page 1: ...Prima SP2 Anaesthetic Machine Range User Manual Partnership for Life WWW ELM FAN IR To download more items go to ELM FAN IR...
Page 2: ...WWW ELM FAN IR To download more items go to ELM FAN IR...
Page 3: ...trained engineers For any enquiry regarding the servicing or repair of this machine contact the nearest accredited Penlon agent or communicate directly with Technical Support Penlon Limited Abingdon...
Page 4: ...re total patient safety Anaesthesia system monitors and patient monitors are very desirable aids for the anaesthetist but are not true clinical monitors as the condition of the patient is also depende...
Page 5: ...rber 13 3 12 AV S Ventilator 13 4 SPECIFICATION 14 4 1 Physical Dimensions 14 4 2 Gas Supplies 15 4 3 Flowmeters 15 4 4 Gas Pressures 16 4 5 Auxiliary Gas Outlets 16 4 6 Oxygen Failure Warning Devices...
Page 6: ...Gas Scavenge System AGSS 28 5 9 Ventilator 29 5 10 Alarm System Testing 30 6 USER MAINTENANCE 6 1 Cleaning and Sterilisation 31 6 2 A200SP Absorber 32 6 3 Service Schedule 33 APPENDIX Disposal at end...
Page 7: ...it is recommended that a request for service advice be made to the nearest Penlon accredited agent This device and any of its constituent parts must be repaired only in accordance with written instru...
Page 8: ...ed resulting in excessive and uncalibrated volumes of anaesthetic drug entering the breathing system Do not install or connect any vaporizers of any description between the Common Gas Outlet CGO and t...
Page 9: ...op shelf must be securely attached Take care when moving a fully loaded machine particularly when negotiating ramps Check that hoses or power leads are not trailing on the floor 3 WARNINGS AND CAUTION...
Page 10: ...selected the machines can be used in open semi open semi closed or closed circuit configurations The range has been designed to give a wide choice of configurations and accessories including Gas supp...
Page 11: ...d the yokes are designed so that the retaining latch cannot be closed unless the index pins are fully engaged 5 3 DESCRIPTION Pipeline Inlets 1 Up to three rear mounted pipeline gas inlets can be fitt...
Page 12: ...pressure source Pressure gauge Pressure regulator Pressure relief valve Audible alarm Restrictor Gas cut off valve normally open Reservoir Pneumatic on off switch Flow control valve variable Filter F...
Page 13: ...pressure relief valve is mounted between the vaporizer back bar and the common gas outlet CGO on the inside face of the machine right hand upright It is designed to prevent fresh gas being delivered t...
Page 14: ...xide valve and is actuated only by movement of the oxygen control The nitrous oxide control knob on the flowmeter module operates a secondary needle valve in the nitrous oxide flow It is positioned do...
Page 15: ...ront panel controls the supply of oxygen and must be in the ON position for normal operation of the machine Flow control of each gas is achieved by a needle valve comprising a polished stainless steel...
Page 16: ...station manifolds are available with each station fitted with two valve capsule assemblies for vaporizer connector block attachment When a vaporizer is installed on a station the valves on that statio...
Page 17: ...the live and neutral wire 3 9 2 Auxiliary Power Supply Sockets CAUTION The requirements of IEC 60601 1 1 apply to any device connected to the auxiliary sockets Users must be aware of the risks of incr...
Page 18: ...ly charged condition Refer also to the user instruction manual supplied with the ventilator Battery charging takes place automatically when the ventilator mains lead is connected to a live mains suppl...
Page 19: ...installation and operation 3 12 AV S Ventilator Interface to Prima SP 2 An interface cable links the gas delivery switch on the anaesthetic machine via a socket 1 on the rear panel of the machine to t...
Page 20: ...ading 10 kg 22 lb evenly distributed Castors Diameter 127 mm 5 inches Front pair braked Ventilator bellows post Bushed to accept 25 4 mm 1 inch or 22 mm 7 8 inch poles Loading 30 kg 66 lb Gas scavengi...
Page 21: ...ide 6 mm Helium 4 mm Air 5 mm Xenon 4 mm Mixed gas 10 mm 4 3 Flowmeters Flow ranges Single flow tubes Oxygen 0 10 L min Nitrous Oxide 0 10 L min Air 0 10 L min Helium 0 10 L min Cascade flow tubes Oxy...
Page 22: ...4 5 Auxiliary Gas Outlets Pipeline supply Gas is supplied at pipeline supply pressure see above Cylinder supply Gas is supplied at reduced pressure from cylinder regulator see above Oxygen Two self s...
Page 23: ...cessary Please refer to section 6 1 4 10 Electrical Supply Power Input NOTE The internal power circuit feeds the machine lighting systems only Option 1 Internal power auxiliary power sockets US CSA sp...
Page 24: ...ments of EN 60601 1 2 electromagnetic compatibility requirements and tests 4 11 Device Classification and Labelling Type B Applied Part Degree of protection against electric shock This symbol denotes...
Page 25: ...be carried out before clinical use WARNING Pre use checks must be performed before each period of clinical use These checks must be supplemented by periodic Function Testing and full Service Testing b...
Page 26: ...n gas delivery switch and note special operating system Check functioning of flowmeters Check function of Mechanical AHD 7 Check correct connection and functioning of the vaporizers 8 Check functionin...
Page 27: ...se Check the gauge reading Check for a flow of air when the flowmeter needle valve is operated 5 2 2 Gas Cylinder Supplies CAUTION Open the cylinder valves slowly to avoid damage to the pressure reduc...
Page 28: ...by gentle rotation of the oxygen knob Check also that the nitrous oxide float reseats on the bottom stop and that the oxygen basal flow continues to flow 5 Operate the other flowmeter control knobs in...
Page 29: ...rizer onto the manifold and lock the vaporizer into position by clockwise rotation of the locking lever through 90o NOTE Do not use excessive force to lock the vaporizer onto the manifold Damage to th...
Page 30: ...upply switch 2 to ON Check for correct function of each auxiliary power outlet 3 Check all electrical equipment including devices powered by the auxiliary power outlets on the rear of the machine 4 Ma...
Page 31: ...naesthetic machine Check all connections for gas tightness 5 5 4 A200SP Absorber Always follow the pre use check procedures given in the instruction manual supplied with the absorber The use of an oxy...
Page 32: ...y 12 Connectors sensor pressure monitor 13 Expiratory Valve Absorber 14 Inspiratory Valve Absorber 15 Inlet from Ventilator Bellows 16 Connector Reservoir Bag 17 Inlet Absorber Fresh Gas Supply 18 Dri...
Page 33: ...system components do not constitute part of the machine but connections between the machine and breathing system should be verified as follows 1 Fit a patient circuit to the inspiratory connector 1 a...
Page 34: ...lent to and in some cases exceeds the requirements for system leakage rates quoted in national standards worldwide 5 7 Oxygen Flush Check for a high flow of oxygen through the CGO outlet when the flus...
Page 35: ...ent of the interface cable at the socket 1 on the rear panel of the Prima SP2 and the connection on the AV S control unit marked Prima master Check the interface system function a Turn the Gas Deliver...
Page 36: ...ction of the internal gas cut off device is performed as follows 1 Connect oxygen nitrous oxide and air supplies 2 Set the Gas Delivery switch 2 to ON and check that the warning whistle sounds briefly...
Page 37: ...on maintenance and service requirements for the ancillary equipment used with the anaesthetic machine Vaporizers Ventilator Absorber AGSS Monitoring 6 1 Cleaning and Sterilisation CAUTION a Check that...
Page 38: ...ions on draining the canister and disposing of absorbent 1 Set the brakes on the anaesthetic machine front castors 2 Refer to the A200SP user manual and disconnect all cable connectors and hoses inclu...
Page 39: ...ne models On vaporizers with interlock the system must be tested during the vaporizer calibration test 6 3 3 Monitoring Systems Follow the service recommendations detailed in the monitor user instruct...
Page 40: ...t There are no risks associated with disposal of this product Do not dispose of in landfill refer to an approved recycling facility Follow your hospital local state and federal regulations APPENDIX 34...
Page 41: ...ssories approved by Penlon Ltd UK Sales Tel 01235 547036 Fax 01235 547023 E mail uksales penlon co uk International Sales Tel 44 1235 547001 Fax 44 1235 547021 E mail export penlon co uk 35 APPENDIX W...
Page 42: ...36 WWW ELM FAN IR To download more items go to ELM FAN IR...
Page 43: ...WWW ELM FAN IR To download more items go to ELM FAN IR...
Page 44: ...ngdon OX14 3PH UK Technical Support Tel 44 0 1235 547076 Fax 44 0 1235 547062 E mail technicalsupport penlon co uk UK Sales Tel 01235 547036 Fax 01235 547023 E mail uksales penlon co uk International...