PressureMAT™ 

System User Guide 

 

Model PMAT1 

 Model 

PMAT2 

 

Model PMAT3 

Model PMAT4A 

Model PMAT4R   

 
 

Revision 1 

 
 

 

 

 

 
 

 
 

 

www.pendotech.com 

 
 

Summary of Contents for PressureMAT PMAT1

Page 1: ...PressureMAT System User Guide Model PMAT1 Model PMAT2 Model PMAT3 Model PMAT4A Model PMAT4R Revision 1 www pendotech com...

Page 2: ...rsonal injury death or substantial property damage if the warning is ignored Note Note is used to notify the user of installation or operation information which is important but not hazard related WAR...

Page 3: ...ications 9 1 2 3 Software Details 10 2 Using the System 20 2 1 System Setup 20 2 2 Using the System 21 3 Cable Information 22 3 1 Cable for Pressure Inputs 22 APPENDIX A PRODUCT WARRANTY 23 APPENDIX B...

Page 4: ...rmaceutical production processes where the system can be used The four available models are as follows Model Number Number of Inputs Number of Outputs Outputs PMAT1 1 2 1 Relay 1 Analog PMAT2 2 4 2 Re...

Page 5: ...ss the optional alarm latching function is turned on which requires STOP to be pressure to clear the alarm The alarm functions are always active for all channels 1 2 Control System Details 1 2 1 Hardw...

Page 6: ...PressureMAT User Guide Revision 1 Page 6 29 BACK PANEL CONFIGURATIONS Two Input Unit if two relays for each channel 4 20mA will be labeled OUTPUT RELAY Three Input Unit...

Page 7: ...PressureMAT User Guide Revision 1 Page 7 29 Four Input Unit 4R shown if 4A OUTPUT RELAY is labeled OUTPUT 4 20mA...

Page 8: ...to CLOSED position with alarm condition b NORMALLY CLOSED wired to terminals R and T via supplied connector will switch to OPEN position with alarm condition c POWER REMOVED OPEN S and T position is...

Page 9: ...e Sensor Inputs Configured for PendoTECH Single Use Pressure Sensors See Appendix B Connector on reusable cable DB15 Excitation voltage 4 096V 0 1 Relay Output Specifications for relay used for the al...

Page 10: ...re is no power switch so the system cannot be accidentally turned off When the wall power supply is connected to the system and plugged into a wall outlet the welcome screen will appear momentarily fo...

Page 11: ...e cursor left UP ARROW Three Purposes 1 When on the HOME screen used to select different input channels 2 In the program menus used to select different sub menus 3 Within a program menu used to change...

Page 12: ...tact relay associated with that channel will switch i e if wired as normally CLOSED it will OPEN b The display screen will blink ALARM on the channel with the alarm c An audible tone for 30 seconds du...

Page 13: ...e SELECT PROG button will escape from that program menu without saving changes made before confirming the entry Channel X Input Viewing Pressure 0 0 psi Service Time XXXX hrs SELECT home display value...

Page 14: ...the DOWN ARROW UP ARROW buttons on the keypad to change the values Note the location of the minus sign is the furthest to the left and the easiest way to access this is by scrolling to the right until...

Page 15: ...at the frequency programmed in the Global Settings Menu contact PendoTECH for more details on data collection from the PressureMAT Press the SELECT PROG with Comm Port selected The port settings are...

Page 16: ...e pressure value is displayed for the selected channel if Name is selected on the Output Viewing menu A wide range of alpha numeric digits can be selected Use the RIGHT ARROW button on the keypad unti...

Page 17: ...ART can be pressed at any time to return to the HOME screen Information If SELECT PROG button pressed system information including firmware version is displayed Press the BACK button to return to the...

Page 18: ...xt Report DDMMMYY HH MM SS Report Start DDMMMYY HH MM SS Report Rate 000 hrs Daylight Save Off This menu is used mainly for controlling data output to a PC Note See Appendix C on Data Collection to a...

Page 19: ...press the BACK button and the main channel menu will appear The report start time must be set later than the current time for data output to commence Report Rate Report rate is used to set the frequen...

Page 20: ...ected to the system 4 Connect system to appropriate power source Note If a pressure sensor is not connected when power is supplied to the system the pressure will go to a value greater than 75 psi 5 1...

Page 21: ...for 30 seconds c The associated relay will switch When an alarm condition goes away all indicators automatically go back to normal unless the optional alarm latching function is turned on which requi...

Page 22: ...s 12 feet 3 7 m in length with a 4 pin connector to connect a pressure sensor on one end and a male DB 15connector to connect to the control system on the other end DB15 connector connected to PendoTE...

Page 23: ...anty and shall be at Buyer s expense Seller shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized...

Page 24: ...vered by a separate product warranty from the PressureMAT system General Information They must be qualified by the user for suitability for an application WARNING EACH PROSPECTIVE USER MUST TEST THE S...

Page 25: ...ivate Normal Mode but in the future open the PressureMAT4 sw3 file by double clicking this file is available from PendoTECH 11 Using the Input Programming menu in the PressureMAT select the Comm Port...

Page 26: ...PressureMAT User Guide Revision 1 Page 26 29 APPENDIX D Panel Mount of System All dimensions in inches 0 015 PressureMAT...

Page 27: ...s of the following EC Directive s when installed in accordance with the instructions contained in supplied product documentation 89 336 EEC EMC Directive as amended by 92 31 EEC and 93 68 EEC 73 23 EE...

Page 28: ...l ABS Plastic top and bottom pieces of box sealed with gasket seal rated to IP66 Note Hole where is valve is mounted is not water tight Pinch Valves BioChem Valve Series 100P http www bio chemvalve co...

Page 29: ...PressureMAT User Guide Revision 1 Page 29 29...