PDI-772HE
10 Tap AC Hospital-grade
Central Power Supply
Document Number:PD196-038R2
Page 1: ...PDI 772HE 10 Tap AC Hospital grade Central Power Supply Document Number PD196 038R2...
Page 2: ...ant to understand that these CAUTIONS and WARNINGS are not exhaustive PDI could not possibly know evaluate and advise the service industry of all conceivable methods in which service might be done or...
Page 3: ...it breaker if tripped If resetting the output circuit breaker does not clear the fault remove the central power supply from service An internal fault has occurred that requires bench repair Individual...
Page 4: ...t With a know RF input signal level to the central power supply RF IN check for an insertion loss of 14 dbmv at each F output Any insertion loss greater than 14 dbmv indicates a fault Remove from serv...
Page 5: ...e cord 2 Remove four screws from the access cover 3 Disconnect all RF output coax cables Close access cover and secure with screws 4 Loosen and remove the four power supply mounting bolts 5 Remove the...