22
1
2
3
4
7.
CAMWARE 4 SOFTWARE
This chapter contains only camera-specific additions to the
pco.camware manual. All main functions and explanations can be
found in the pco.camware manual.
7.1
HARDWARE IO CONTROL
Change settings using the drop-down menu.
For electrical specification, see chapter
Exposure Trigger
If checked on, a signal for
External Exposure Start
or
External
Exposure Control Trigger Mode
(see chapter
6.3.1
) is
accepted at the
Exposure Trigger
SMA input #1.
Exposure Trigger:
On; Off
Signal Polarity:
Rising; Falling
Aquire
Enable
If checked on, a signal for Acquire Mode or Sequence Trigger Mode
is accepted at the
Acquire Enable
SMA input #2.
Acquire Enable:
On; Off
Signal Polarity:
High; Low
Status Busy
If checked on, a signal indicating busy status is provided at the status
busy output. Once an acceptable trigger edge is received, ‘busy’
goes to status high. As soon as ‘busy’ goes low again, a new trigger
edge is accepted.
Status Busy:
On; Off
Signal Polarity:
High; Low
Status Expos
If checked on, a signal indicating exposure or line status is provided
the status output.
Status Expos
indicates the actual exposure
window for one frame.
Status Line:
use the falling line edge for
optimized synchronization (minimized jitter; see page
Select IO Signal:
Status Expos; Status Line
Signal timing:
Show time of ‘First Line’; Show common time of ‘All
lines’; Show time of ‘Last line’; Show overall time of ‘All lines’
Status Expos:
On; Off
Signal Polarity:
High; Low
1
2
3
4
1
2
3
4