![PCE Instruments PCE-MA 100 Manual Download Page 11](http://html1.mh-extra.com/html/pce-instruments/pce-ma-100/pce-ma-100_manual_1516106011.webp)
Manual
11
5.2 Drying modes
Press the
key to get to the drying mode selection screen. Here you can select different drying modes
and set up new ones.
A number which indicates the memory location flashes on the left-hand side of the display. The user can
retrieve up to 16 different memory locations, using the arrow keys
and
. The saved drying
parameters are displayed along with the selected memory location.
If you wish to use the drying parameters of a certain memory location, press the
key. The moisture
balance will then take over the drying parameters and you can start the measurement.
If you want to save some new drying parameters to a memory location, select the memory location and
confirm with the
key.
5.2.1
Specify heat-up mode
After the selection of the memory location, the drop icon will flash in the upper side of the display.
The user can select one out of three heat-up speeds.
Standard heat-up mode
This is the default mode which is suitable for most sample types.
In this heat-up mode, 120 °C are reached
after approx. 4 minutes
.
Quick heat-up mode
This mode is suitable for samples with a high moisture content.
In this heat-up mode, 120 °C are reached
after approx. 1 minute
.
ATTENTION:
When using this heat-up mode, there might be increased temperatures in the
drying chamber for a short period as the sensor and the control need some time to regulate the
temperature.
Slow heat-up mode
This mode is suitable for samples with a low moisture content.
In this heat-up mode, 120 °C are reached
after approx. 8 minutes
.
ATTENTION:
When using this heat-up mode, it is particularly important that the drying time of the
sample is adapted to the drying mode as otherwise the drying process could be terminated (time
expired) although there is still some residual moisture in the sample.
You can select the desired heat-up mode using the arrow keys
and
.
To confirm the selected heat-up mode, press the
key.