background image

 

 

©  PCE Instruments 

11 

 

7.7.2 

Long-term memory 

The long-term memory contains all the measured values that were saved there  from the short-
term memory via the Save option. 
To call up the stored measured values, proceed as described in the menu. 
Navigate to the desired view as described in chapter 7.3 
24 measured values are listed per page. Press the 

"② 

Down" key to call up the other pages and 

the graphic display. Press the 

"① 

Up" key to scroll back one page. 

In the list, the min and max values are marked by red numbers.  
The graph shows a bar chart with the limit value setting. The measured values that  are outside 
the set limit values are shown with a red bar. 
In the header, the unit of measurement, the average value and the number of measurements are 
shown. 
Press the key 

"③ 

On / Off, Menu" to return to measuring mode. 

7.8 

Calibration 

In order to ensure the accuracy of the measurement, it is necessary to calibrate the meter. 
Before each measurement, carry out a control measurement on the calibration standard without 
and with a foil of defined thickness. If there are deviations in the accuracy from the technical data, 
carry out at least one zero and one offset calibration. After resetting the meter to factory settings 
or when the system has not been used for more than 10 days, a base calibration should also be 
performed. 
Always  use  the  calibration  standard  corresponding  to  the  base  material  of  your  measurement 
object for calibration. 

7.8.1 

Base calibration 

For a base calibration, prepare a calibration standard and the calibration foils with increasing 
thicknesses. 
Navigate to the "Base calibration" menu option as described in chapters 7.2and 7.3 and open it.  
Follow the instructions on the display and start measuring on the calibration standard without a 
foil. You will then be prompted to perform one measurement after the other on a film with a 
thickness of approx. 52, 105, 262, 490, 991 and 1486 µm. A beep means that the next 
measurement is due. 
Always carry out the respective measurement on the film that is closest to the thickness 
specified by the coating thickness gauge. 
The set of foils included in the scope of delivery is matched to the device. The set of foils may 
show slight deviations from the foil thicknesses required by the test instrument during base 
calibration. These can be neglected during the base calibration. 

7.8.2 

Offset calibration 

For calibration of the measured values, prepare a calibration standard and the calibration foils. 
Navigate to the menu option "offset calibration" as 7.3in chapters 7.2 and 7.3and open it. 
Place  the  calibration  foil  on  the  calibration  standard  that  most  closely  matches  the  coating 
thickness on the object to be measured. 
Place the meter on the film in the same way as for a measurement. The gauge will perform a few 
measurements. Correct the displayed value with the keys 

"① 

Up" or 

"② 

Down" and confirm the 

setting with the 

"⑤ 

OK" key. 

 

 

Summary of Contents for PCE-CT 24FN

Page 1: ...ng Thickness Gauge User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last cha...

Page 2: ...e 2 4 System description 3 5 Preparation 4 6 Battery indicator 4 7 Operation 4 7 1 On off 4 7 2 Menu settings 4 7 3 Menu levels 5 7 4 Measurement 6 7 5 Display indication 7 7 6 Rotate display 10 7 7 M...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...0 5 mm Memory 100 groups of 15 readings each Display 2 25 inch LC display Ambient temperature 0 50 C 32 122 F Power supply batteries 3 x 1 2 V AA Ni MH 800 mAh Power supply mains adaptor Primary 100 2...

Page 5: ...PCE Instruments 3 4 System description Up Display Down USB socket On Off Menu Battery compartment Back Sensor OK Enter...

Page 6: ...Menu key When switching on make sure that the probe is at least 10 cm away from the object to be measured To switch off the meter press and hold the On Off Menu key for approx 3 seconds Automatic pow...

Page 7: ...s Adjustment of the limit values High Limit Low Limit Limit sound On Off 7 3 4 Memory Memory Look back short term memory Look Data Save Data Clear data Browse from the first group data memory sorted b...

Page 8: ...uccessful measurement is acknowledged by a sound and the measured value is shown on the display The V grooves on the probe help you to align the device on convex measurement objects 7 4 1 Single conti...

Page 9: ...plays with the Up or Down keys Typical display Header Display mode Battery indicator Upper part Number of current measurements Base material Fe magnetic nFe non magnetic Measuring mode SNG Single meas...

Page 10: ...Display mode Battery indicator Upper part Number of current measurements Base material Fe magnetic nFe non magnetic Measuring mode SNG Single measurement CTN Continuous measurement Middle part Current...

Page 11: ...few measurements Max Maximum value of the last few measurements Min Minimum value of the last few measurements High Upper limit value Low Lower limit value Sdev Deviation CV Percentage deviation Sdev...

Page 12: ...1 Short term memory The short term memory contains the values measured after the last switch on When the meter is switched off the measured values not saved to the long term memory are lost To view sa...

Page 13: ...ion standard corresponding to the base material of your measurement object for calibration 7 8 1 Base calibration For a base calibration prepare a calibration standard and the calibration foils with i...

Page 14: ...ur warranty terms in our General Business Terms which you can find here https www pce instruments com english terms 9 Disposal For the disposal of batteries in the EU the 2006 66 EC directive of the E...

Page 15: ...tanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 0583 975 114 Tel 0212 471 11 47 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 Faks 0212 705 53 93 info pce instruments co uk...

Reviews: