9.1 

Zero point calibration  

To  perform  a  zero  point  calibration,  press  the 

 key.  Now  place  the  sensor  on  the 

corresponding zero reference. When "Lift probe up" appears on the display, remove the sensor 

from the reference. You will then be asked if you want to accept the calibration. Press the 

 

key to save the change. To discard the setting, press the 

 key.  

9.2 

One-point calibration  

To perform a 1-point calibration, first go to the calibration menu and select "1-point calib". Now 
place  the  sensor  on  the  corresponding  zero  reference.  When  "Lift  probe  up"  appears  on  the 
display,  remove  the  sensor  from  the  zero  reference.  Now  place  a  foil  reference  on  the  zero 
reference.  Now carry  out  a measurement  of the  reference  by  positioning  the  probe  on  the  foil. 
When "Lift probe up" appears on the display, remove the sensor from the reference. The display 
will  now  show  the  measured  value.  With  the  arrow  keys,  you  can  now  set  this  value  to  the 

calibration reference. Confirm the set value with the 

 key. You are then asked whether you 

want to accept the calibration. Press the 

 key to save the change. To discard the setting, press 

the 

 key. 

9.3 

Five-point calibration 

To perform a five-point calibration, first go to the calibration menu and select "5-point calib". Now 
place  the  sensor  on  the  corresponding  zero  reference.  When  "Lift  probe  up"  appears  on  the 
display,  remove  the  sensor  from  the  zero  reference.  Now  place  a  foil  reference  on  the  zero 
reference. Now carry out a measurement of the reference by positioning the probe on the film. 
When "Lift probe up" appears on the display, remove the sensor from the reference. The display 
will now show the measured value. With the arrow keys, you can now set this value to the value 

of the calibration reference. Confirm the set value with the 

 key. The second calibration point 

will follow now. Now use another calibration reference and repeat the calibration procedure until 
you  reach  the  last  calibration  point.  Then  you  are  asked  whether  you  want  to  accept  the 

calibration. Press the 

 key to save the change. To discard the setting, press the 

 key. 

Important: 

When instructed to lift the probe, lift the probe quickly and place it 30 cm away from 

the reference before changing the reference film. 
 

 

Summary of Contents for PCE-CT 21BT

Page 1: ...Coating Thickness Gauge User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Las...

Page 2: ...off 4 7 Measurement 4 7 1 Continuous measurement 4 8 Menu 4 8 1 Menu structure 5 9 Calibration 5 9 1 Zero point calibration 6 9 2 One point calibration 6 9 3 Five point calibration 6 10 Data group 7...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...Fe Fe NFe Fe NFe Probe external internal external Measurement range 0 1500 m Resolution 0 1 m within meas range 0 99 9 m 1 m within meas range 100 1500 m Accuracy 1 m 2 of the coating thickness Units...

Page 5: ...terial m Unit Adjustable units m and mil v Limit value Is displayed when the set limit value is exceeded or fallen below 17 Number of measurements Number of measured values 4 1 Key description Symbol...

Page 6: ...ccur To carry out further measurements place the sensor on the next measuring spot Important In order to obtain the best results it is important to take measurements on a smooth and non slip surface D...

Page 7: ...OFF Unit m mil Device information About Sensor type Type Serial number Serial Hardware version Hardware Software version Software 9 Calibration To perform a calibration the meter has various function...

Page 8: ...asked whether you want to accept the calibration Press the key to save the change To discard the setting press the key 9 3 Five point calibration To perform a five point calibration first go to the ca...

Page 9: ...e memory 11 Limit value alarm To set limit values first go to the limit alarm menu Under the item Alarm on you can activate the limit alarm function You can set the limits as follows Upper limit value...

Page 10: ...this is not available the driver may be missing You can find the driver in the program folder under the driver folder If you have a 64 bit system use the installation file CP210xVCPInstaller_x64 If y...

Page 11: ...le device first download the Coatingmaster app Then switch on the Bluetooth function in the measuring device Now connect the coating thickness gauge via the app To do this select Coating_XXXX in the a...

Page 12: ...ngs to the app Setting Meaning Measuring sounds During a measurement a sound can be heard Measuring vibration When a measurement is made the vibration alarm is activated Alarm sounds When the limit is...

Page 13: ...any measurements at the edge Incorrect placement of the probe Always align the probe vertically and quickly with the surface Do not perform a measurement by force The probe must touch the test object...

Page 14: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Page 15: ...tanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 0583 975 114 Tel 0212 471 11 47 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 Faks 0212 705 53 93 info pce instruments co uk...

Reviews: