Manual 

 

17 

8.2 

System setup 

8.2.1 

Language option 

Here you can choose the language of the instrument. You may choose between English, Chinese, 
Portuguese, Turkish and German. Use the arrow keys to choose and the “ENTER” key to confirm.  

8.2.2 

Sound 

Here you can choose what sounds the instruments is supposed to make. Make your choice with the 
arrow keys and confirm with “ENTER”. 

8.2.3 

Battery manager 

You have got two options here. If “ON” is ticked, the instrument switches itself off after 30 minutes of 
inactivity. If the option “OFF” is ticked, the instrument changes to Standby-mode in case of inactivity. 
Make your choice by using the arrow keys and confirm with “ENTER”.  

8.2.4 

Display 

Here the contrast of the display can be set. Use the right / left arrow keys to move the slide control. 
Confirm your choice with “ENTER”. 

8.2.5 

Date 

You can set time and date here. The following settings can be made: year, month, day, hour, minute and 
weekday. Choose the correct field with the arrow-up and arrow-down keys and put in the correct settings, 
using the left and right arrow keys. Confirm your entered data using the “ENTER” key.  

8.2.6 

Restore default 

Here you can restore the factory settings. Please enter the password (standard: ”888888”) and press 
“ENTER”. Then press “ENTER” again to confirm the reset. Press the “MENU” key to cancel.  
Note: All data is going to be deleted. This includes calibration details, settings and measured data.  

8.3 

Save setup 

8.3.1 

Autosave 

Here the automatic storage function can be activated or deactivated. You one option and confirm by 
pressing “ENTER”. 

8.3.2 

Result display 

Here you can review the saved data. The values are assorted by their numbers. With the up / down keys 
you can scroll through the values. By pressing “ENTER” you can open one slot to get detailed 
information.  
If you need information on one specific slot number, press “SCALE”. A search window will appear. Insert 
the number you are looking for using the arrow keys and confirm your choice with “ENTER”.  

8.3.3 

On-line operation 

Here you can activate the data transmission to a PC. Choose an option by using the arrow keys and 
confirm with “ENTER”.  

8.3.4 

Part deletion 

Here you can choose and delete the storage slots. Use the arrow keys to choose the storage slot and 
press “ENTER” to choose one slot. If you have chosen all slots, press “SCALE” and confirm by choosing 
“YES” and pressing “ENTER” afterwards.  

8.3.5 

All deletion 

Here you can delete all measured data. Press “ENTER” to confirm and “MENU” to cancel. 
Note: This procedure may take some minutes.  

Summary of Contents for PCE-5000

Page 1: ...er FL 33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business...

Page 2: ...System setup 17 8 2 1 Language option 17 8 2 2 Sound 17 8 2 3 Battery manager 17 8 2 4 Display 17 8 2 5 Date 17 8 2 6 Restore default 17 8 3 Save setup 17 8 3 1 Autosave 17 8 3 2 Result display 17 8...

Page 3: ...ness tests of central axes thin walled pipes and vessels Hardness testing of tires and turbine rotor Hardness tests of cutting blades Hardness testing of welded parts Hardness tests of certain opening...

Page 4: ...the recharger with wet or damp hands otherwise there is a danger of an electric shock Declaration Without the prior written consent of the Company and subsidiaries for all products that are referring...

Page 5: ...ications Measurement range Rockwell 20 3 68 0 HRC 41 0 100 0 HRB 61 0 85 6 HRA Brinell 85 650 HB Vickers 50 999 HV Tensile strength 255 2180 N mm Accuracy Rockwell HRD 1 5 HR Brinell HB 3 Vickers HV 3...

Page 6: ...ilicone 1 x RS 232 interface cable 1 x robust carry case 1 x user manual 4 1 Optional accessories Probes There are different probes depending on the testing forces available 10 N 50 N 98 N Supporting...

Page 7: ...ystem description 5 1 Main structure and basic principle Front and back view of the ultrasonic hardness tester 1 2 6 3 4 5 1 4 pin socket 2 8 pin socket 3 Power button 4 USB interface 5 Operation pane...

Page 8: ...Enter In measurement mode start measurement In menu back Menu During measurement open menu In menu back Power on off brightness Press shortly switch on off backlight Press long switch on off the devi...

Page 9: ...can see how many measurements have to be made in the current test series In addition to that you can see in which slot the measured data has been stored Material information Here you can see which mat...

Page 10: ...Manual 10 5 4 Measurement probe 8 pin socket Vickers diamond Note The standard probe is preset to 20 N Other probes are available as optional accessory...

Page 11: ...300 16 Material hardness HV Diagonal line of the impression of the inspection stamp 800 68 600 79 300 111 Note The table only shows the values for the 20 N probe Minimum thickness of the tested objec...

Page 12: ...ensions and weight of the part to be tested From the above analysis all probes request coating layer or surface layer less than 1 mm The hardness value will change if there is resonance The highest va...

Page 13: ...er right corner from the main menu to remind you to charge the battery in time Plug one end of the charger into the left socket of the instrument and insert the plug into a 230 V outlet and charge the...

Page 14: ...ar a clicking sound the connections are correct 7 2 Measurement To perform a measurement follow these steps 1 Press Power and move the ON OFF switch to ON so that the power supply is turned on You wil...

Page 15: ...settings you need to stick to that setting Otherwise the measurement is not completed The ultrasonic hardness meter is a highly precise instrument Take care to always hold the probe with two hands in...

Page 16: ...an rename the calibrations Use arrow and ENTER keys to confirm the entered data Note When in measurement mode you can also get into this mode by pressing the MAT key Hardness scale Here you can choose...

Page 17: ...re the factory settings Please enter the password standard 888888 and press ENTER Then press ENTER again to confirm the reset Press the MENU key to cancel Note All data is going to be deleted This inc...

Page 18: ...ibration is necessary 9 3 Calibration procedure 9 3 1 Calibration with two calibration standards Press MENU in the main menu Move to test setup and choose the Calibration setup Insert the password 888...

Page 19: ...calibration settings you can also enter the tolerance thresholds If you have entered the upper and lowest value the instrument shows if the measured value is within that area every time you measure If...

Page 20: ...use them or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 11 Contact If you have any questions about our rang...

Reviews: