background image

 

Model 394C06

 

Platinum Stock Product Handheld shaker, 1g at 159.2 Hz (for up to 210 grams total weight of

sensor, cable and mtg. accessories).

 

Installation and Operating Manual

 

For assistance with the operation of this product,

contact the PCB Piezotronics, Inc.

 

Toll-free: 716-684-0001

24-hour SensorLine: 716-684-0001

Fax: 716-684-0987

E-mail: [email protected]

Web: www.pcb.com

Summary of Contents for 394C06

Page 1: ...0 grams total weight of sensor cable and mtg accessories Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...ibration Shaker Operating Guide with Enclosed Warranty Information 3425 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684 3823 Toll Free Line 1 800 828 8840 MANUAL NUMBER 18295 MANUAL REVISION B ECO 51616 ...

Page 7: ...MBER 18295 MANUAL REVISION B Table of Contents Introduction page 3 Description page 3 Operation page 4 Warnings Cautions page 6 Warranty page 7 Service page 7 Return Procedure page 7 Total Customer Satisfaction page 8 ...

Page 8: ...ed for velocity and displacement at 10 mms 1 RMS or 10 mms 1 peak and 10 µm RMS or 10 µm peak to peak respectively To change between RMS and peak depress the corresponding side of the rocker switch in the options cavity of the shaker For example to choose RMS depress the left side of the rocker switch as illustrated in Figure 1 Model 394C06 also provides quick single sensor validation in complex m...

Page 9: ...gure 1 The unit will turn on and internal power to the shaker will be ramped up providing a smooth gradual increase in vibration amplitude This process prevents shock induced overloads to the test sensor A special reference sensor within the unit constantly measures the g level and internal controls adjust the shaker magnitude to maintain 1 g output while maintaining the frequency of 159 2 Hz A re...

Page 10: ...Points Calibration of the internal feedback circuit can be verified via test points located in the options cavity see Figure 1 The measured value in volts AC should agree with the calibration certificate This is useful for a quick field check of unit integrity which may be necessary if the shaker is dropped and a reference standard is not immediately available This test is not a guarantee of shake...

Page 11: ...rge ESD can damage electrical devices Similar to a capacitor a cable can hold a charge caused by triboelectric transfer such as that which occurs in the following Laying on and moving across a rug Any movement through air The action of rolling out a cable and or Contact with a non grounded person The PCB solution for product safety Connect the cables only with the AC power off Temporarily short th...

Page 12: ...t quotation will be provided prior to servicing at no charge Before returning the unit please consult a factory PCB applications engineer concerning the situation as certain problems can often be corrected with simple on site procedures Return Procedure To expedite returned instrumentation contact a factory PCB applications engineer for a RETURN MATERIAL AUTHORIZATION RMA NUMBER When requesting th...

Page 13: ...funded PCB offers to all customers at no charge 24 hour phone support This service makes product or application support available to our customers day or night seven days a week When unforeseen problems or emergency situations arise call the PCB Hot Line at 716 684 0003 and an application specialist will assist you 3425 Walden Avenue Depew NY 14043 2495 Phone 716 684 0003 USA Fax 716 684 3823 INTL...

Page 14: ...d In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have identical specifications and accessories as listed for the standard model except where noted below More than one option may be used M Metric Mount Acceleration Output 3 1 02 g rms 10 0 m s rms NOTES 1 ...

Page 15: ...IONS METRIC DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 2 5 FOOT POUNDS 3 7 NEWTON METERS MOUNTING HOLE PREPARATION 218 5 54 X 300 7 62 MIN 1 4 28 UNF 2B X 200 5 08 MIN F M6 X 0 75 M6 X 1 00 M8 X 1 25 MOUNTING INSTRUCTIONS ENGLISH DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 3 7 NEWTON METERS 2 5 FT POUNDS M8 X 1 25 MOUNTING HOLE PREPARATION 6 75 266 X 8 64 340 MIN M8 X 1 25 6H X ...

Page 16: ...LETS AND RADII 003 005 FILLETS AND RADII 0 07 0 13 OUTLINE DRAWING 5746 1 OF 1 625X MODEL 394C06 HAND HELD CALIBRATOR DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 KRM 11 1 18 KRM 11 1 18 CPH 11 1 18 5746 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotron...