PCB Piezotronics 357a100 Installation And Operating Manual Download Page 18

MANUAL: 55308 
REV:  A 
ECO:  49582 

13 

6.1.3  PCB CALIBRATION 

PROCEDURE 

Numerous precautions  are taken at PCB to insure 
accurate  and  repeatable  results.    This  section 
provides  a  brief  overview  of  the  primary  areas  of 
concern. 
 
Since the Back-to-Back Calibration technique relies 
on  each  sensor  experiencing  an  identical 
acceleration  level,  proper  mounting  of  the  test 
sensor  to  the  reference  standard  is  imperative.  
Sensors  with  mounting  holes  are  attached  directly 
to  the  reference  standard  with  a  stud  tightened  to 
the  recommended  mounting  torque.  A  shouldered 
mounting stud is typically used to prevent the stud 
from “bottoming out” in the hole.  
 
Both  mounting  surfaces  are  precision-machined 
and  lapped  to  provide  a  smooth,  flat  interface 
according  to  the  manufacturer’s  specification.    A 
thin layer of silicone grease is placed between the 
mating  surfaces  to  fill  any  imperfections  and 
increase  the  mounting  stiffness.    The  cables  are 
stress-relieved  by  first  routing  them  to  the  shaker 
head,  securing  them  with  tape  or  cable  ties,  then 
routing  them  to a  nearby  stationary  location.    This 
reduces cable motion, which is especially important 
when  testing  charge  output  sensors  and  helps  to 
prevent  extraneous  noise  or  stresses  from  being 
imparted into the system.  A typical set-up is shown 
in Figure 4.  
 

 

 

Figure 4.

 Typical Calibration Set-Up 

 
Adhesively mounted sensors use similar practices.  
However,  in  this  case,  a  small  portion  of  quick-
bonding  gel  or  similar  temporary  adhesive  is  used 
to  attach  the  test  sensor  to  a  reference  standard 
designed with a smooth, flat mounting surface. 

 
In addition to mounting, the selection of the proper 
equipment is critical.   Some of the more important 
considerations  include:  1)  the  reference  standard 
must be specified and previously calibrated over the 
frequency and/or amplitude range of interest; 2) the 
shaker  should  be  selected  to  provide  minimal 
transverse  (lateral)  motion  and  minimal  distortion; 
and  3)  the  quality  of  the  meters,  signal  generator, 
and  other  devices  should  be  selected  so  as  to 
operate within the limits of permissible error. 
 

7.4 COMMON MISTAKES 

Most  calibration  errors  are  caused  by  simply 
overlooking  some  of  the  fundamental  principals  of 
dynamics.  This section attempts to address some 
of the more common concerns. 
 
For  stud-mount  sensors,  always  mount  the 
accelerometer  directly  to  the  reference  standard.  
Ensure that the mounting surfaces are smooth, flat, 
and free of any burrs.  Always use a coupling fluid, 
such as silicone grease, in the mounting interface to 
maintain  a  high  mounting  stiffness.    Mount  the 
sensor 

according 

to 

the 

manufacturer’s 

recommended mounting torque.  DO NOT use any 
intermediate  mounting  adaptors,  as  the  mounted 
resonant  frequency  may  be  reduced  and  thereby 
compromise  the  high-frequency  performance.    If 
necessary, use adaptor studs. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for 357a100

Page 1: ...erometer Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...with PIEZOELECTRIC CHARGE MODE ACCELEROMETERS SPECIFICATION SHEET INSTALLATION DRAWING AND CALIBRATION INFORMATION ENCLOSED PCB ASSUMES NO RESPONSIBILITY FOR DAMAGE CAUSED TO THIS PRODUCT AS A RESULT...

Page 7: ...nector 4 2 5 In Line Differential Charge Amplifier 4 3 INSTALLATION OVERVIEW 4 3 1 Equipment Inspection 4 3 2 Polarity Test 4 3 3 Sensor Location 4 3 4 Mounting Sensor 5 3 5 Route Mineral Insulated Ha...

Page 8: ...re high temperatures preclude the use of accelerometers with built in microelectronics Enclosed is a Specification Sheet which lists the complete performance characteristics of the particular accelero...

Page 9: ...harge Amplifier The differential charge amplifier is purchased separately A conventional method for conditioning the high impedance signal generated by a charge output sensor is to use a differential...

Page 10: ...priate charge amplifier specification sheet for information on low frequency characteristics Figure 1 Assorted Mounting Configurations and Their Effects on High Frequency 3 4 1 STUD MOUNT This mountin...

Page 11: ...est object STEP 1 Prepare a smooth flat mounting surface A minimum surface finish of 63 in 0 00016 mm generally works best STEP 2 Stud mount the sensor to the appropriate adhesive mounting base accord...

Page 12: ...amp Cable To minimize triboelectric motion induced noise from the cable interfering with the sensors high impedance charge output cable clamps must be used Clamp the cable as close to the transducer a...

Page 13: ...ation Once each element is connected allow a few minutes for the system to thermally stabilize Place the switch on the charge amplifier in the OPERATE position and proceed with the measurement It is o...

Page 14: ...vices As demonstrated in Equation 1 the steady state DC output voltage Eo is Equation 1 i f i o R R e E 1 This equation shows that if the input leakage resistance at the charge amplifier is extremely...

Page 15: ...mentioned and one comprised of series capacitor Cs and total equivalent shunt resistance Ri This new cutoff frequency is Equation 3 s i o C R 16 f To avoid compromise of the low frequency response est...

Page 16: ...s checking against other equipment manufacturer recommendation and any risk associated with incorrect readings International standards such as ISO 10012 1 provide insight and suggested methods for det...

Page 17: ...ay be automated and vary depending on the type and number of accelerometers being calibrated Figure 2 Typical Back to Back Calibration System Because the acceleration is the same on both sensors the r...

Page 18: ...4 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the test sensor...

Page 19: ...rd Validate your calibration system with another accelerometer prior to each calibration session Check with the manufacturer for exact system specifications 7 5 CONCLUSIONS Without an adequate underst...

Page 20: ...5 in x 1 88 in 19 mm x 38 1 mm x 47 8 mm Weight 2 32 oz 65 8 gm 1 Electrical Connector 7 16 27 2 Pin 7 16 27 2 Pin Electrical Connection Position Side Side Mounting Through Holes 3 Through Holes 3 5 A...

Page 21: ...NG 63685 1 OF 1 2X ACCELEROMETER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 JDM 2 17 16 ECB 2 17 16 GCD 2 17 16 63685 PCB Piezotronics Inc claims proprietary rights in the information disclosed her...

Reviews: