PCB Piezotronics 357a100 Installation And Operating Manual Download Page 13

MANUAL: 55308 
REV:  A 
ECO:  49582 

 

4  POWERING  

 

4.1  Installation 

Before connecting the low-noise cable from the accelerometer to the charge amplifier, be sure to ground both 
the charge amplifier and the cable.  This ensures that an excessive static charge that may have accumulated 
across the accelerometer or cable is harmlessly discharged.  Failure to observe this precaution can result in the 
destruction of the input FET of certain amplifiers. 
 

 
 
Connect  the  transducer  to  the  input  of  a  PCB 
differential  or  equivalent  charge  amp  using  low 
noise  cable.   

Note:    For  optimum  noise 

performance, the cable length between the sensor 
and the charge AMP should be minimized. 

 
Connect the output of the charge AMP to any ICP 
® signal conditioner using standard cable.  Finally, 
the output of the signal conditioner may then be 
connected to an oscilloscope or other monitoring 
device.  This output will be an AC signal (see 

specification

 for actual frequency response) with 

a DC bias.  Many PCB signal conditioners remove 
the bias via an AC coupling circuit. 

 

4.2  Operation 

Once  each  element  is  connected,  allow  a  few 
minutes for the system to thermally stabilize.  Place 
the switch on the charge amplifier in the OPERATE 
position and proceed with the measurement. 
 
It 

is 

often 

convenient 

to 

normalize 

the 

accelerometer  and  charge  amplifier  system  to  a 
precise sensitivity, such as 10.0  or 100.0 mV/g for 
ease  of  data  analysis.    This  is  accomplished  with 
most  PCB  laboratory  charge  amplifiers  and  some 
miniature in-line units as well. 

 
For fixed sensitivity charge converters, the system 
sensitivity  (mV/g)  is  determined  as  the  product  of 
the  charge  amplifier  sensitivity  (mV/pC)  and  the 
charge sensitivity of the accelerometer (pC/g). 

 
Note:

   

When  using  charge-amplified  systems,  the 

noise floor of the system is dependent on the input 
capacitance to the charge amplifier.  Since the cable 
adds to the capacitance and to minimize the noise 
threshold,  keep  the  cable  length  between  the 
accelerometer  and  the  charge  amplifier  to  a 
minimum.  Cable length does not affect the system 
sensitivity of charge-amplified systems.

 

 
Since charge amplifier operation varies, please 
contact the respective signal conditioner 
manufacturer or check the product manual for 
additional information. 

 
 

HIGH-TEMPERATURE 
 

OPERATION 

 

5.1    Introduction 

 

Summary of Contents for 357a100

Page 1: ...erometer Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...with PIEZOELECTRIC CHARGE MODE ACCELEROMETERS SPECIFICATION SHEET INSTALLATION DRAWING AND CALIBRATION INFORMATION ENCLOSED PCB ASSUMES NO RESPONSIBILITY FOR DAMAGE CAUSED TO THIS PRODUCT AS A RESULT...

Page 7: ...nector 4 2 5 In Line Differential Charge Amplifier 4 3 INSTALLATION OVERVIEW 4 3 1 Equipment Inspection 4 3 2 Polarity Test 4 3 3 Sensor Location 4 3 4 Mounting Sensor 5 3 5 Route Mineral Insulated Ha...

Page 8: ...re high temperatures preclude the use of accelerometers with built in microelectronics Enclosed is a Specification Sheet which lists the complete performance characteristics of the particular accelero...

Page 9: ...harge Amplifier The differential charge amplifier is purchased separately A conventional method for conditioning the high impedance signal generated by a charge output sensor is to use a differential...

Page 10: ...priate charge amplifier specification sheet for information on low frequency characteristics Figure 1 Assorted Mounting Configurations and Their Effects on High Frequency 3 4 1 STUD MOUNT This mountin...

Page 11: ...est object STEP 1 Prepare a smooth flat mounting surface A minimum surface finish of 63 in 0 00016 mm generally works best STEP 2 Stud mount the sensor to the appropriate adhesive mounting base accord...

Page 12: ...amp Cable To minimize triboelectric motion induced noise from the cable interfering with the sensors high impedance charge output cable clamps must be used Clamp the cable as close to the transducer a...

Page 13: ...ation Once each element is connected allow a few minutes for the system to thermally stabilize Place the switch on the charge amplifier in the OPERATE position and proceed with the measurement It is o...

Page 14: ...vices As demonstrated in Equation 1 the steady state DC output voltage Eo is Equation 1 i f i o R R e E 1 This equation shows that if the input leakage resistance at the charge amplifier is extremely...

Page 15: ...mentioned and one comprised of series capacitor Cs and total equivalent shunt resistance Ri This new cutoff frequency is Equation 3 s i o C R 16 f To avoid compromise of the low frequency response est...

Page 16: ...s checking against other equipment manufacturer recommendation and any risk associated with incorrect readings International standards such as ISO 10012 1 provide insight and suggested methods for det...

Page 17: ...ay be automated and vary depending on the type and number of accelerometers being calibrated Figure 2 Typical Back to Back Calibration System Because the acceleration is the same on both sensors the r...

Page 18: ...4 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the test sensor...

Page 19: ...rd Validate your calibration system with another accelerometer prior to each calibration session Check with the manufacturer for exact system specifications 7 5 CONCLUSIONS Without an adequate underst...

Page 20: ...5 in x 1 88 in 19 mm x 38 1 mm x 47 8 mm Weight 2 32 oz 65 8 gm 1 Electrical Connector 7 16 27 2 Pin 7 16 27 2 Pin Electrical Connection Position Side Side Mounting Through Holes 3 Through Holes 3 5 A...

Page 21: ...NG 63685 1 OF 1 2X ACCELEROMETER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 JDM 2 17 16 ECB 2 17 16 GCD 2 17 16 63685 PCB Piezotronics Inc claims proprietary rights in the information disclosed her...

Reviews: