PCB Piezotronics 118A07 Installation And Operating Manual Download Page 1

Model 118A07

Charge Output Pressure Sensor

Installation and Operating Manual

For assistance with the operation of this product,

contact PCB Piezotronics, Inc.

Toll-free: 800-828-8840

24-hour SensorLine: 716-684-0001

Fax: 716-684-0987

E-mail: [email protected]

Web: www.pcb.com

Summary of Contents for 118A07

Page 1: ...ure Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Page 2: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Page 3: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Page 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 6: ...rs The short rigid quartz element and stiff diaphragm give the unit a 250 kHz natural frequency and good linearity even below 1 000 psi 3 0 INSTALLATION When choosing an installation method the advant...

Page 7: ...it free from tool chatter marks NOTE It is important that this surface be perfectly smooth and free from nicks or other discontinuities that could cause leaks at high pressures 3 3 Recess Mount The r...

Page 8: ...be thermally insulated with DC 4 silicone grease or its equivalent In place of the silicone rubber one layer of black vinyl electrical tape on the diaphragm may provide sufficient insulation 3 6 Cable...

Page 9: ...ess the ground button of the charge amplifier and adjust electrical zero if necessary Range the amplifier to give the necessary full scale voltage For normal drift free operation switch the charge amp...

Page 10: ...Jack 10 32 Coaxial Jack Weight 0 52 oz 14 6 gm 5 All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right tochange spe...

Page 11: ...001 6 3 0 43 11 0 91 23 1 1 65 41 9 035 040 1 02 89 3 8 24 UNF 2A 312 7 92 HEX 10 32 UNF 2A CO AXIAL ELECTRICAL CONNECTOR 236 237 5 99 6 02 800 20 32 3 8 24 UNF 2B 470 11 94 DRILL AND REAM 2520 0004...

Reviews: