A page will be displayed and will include the request to confirm the execution of the
measurement. Pressing the Accept button the measurement will be performed, by
pressing the Exit button the measurement will not be performed and you will return to the
systems main menu.
The execution of a single measure, unlike the execution of a session, does not require the
instrument to memorize the result, therefore the measured electrical parameters will only
be shown on the display and must be used by the operator to copy them if necessary.
If you want to perform individual measurements and store the result on the system, it is
advisable to create a "Generic" session and perform the measurements, possibly taking
note of the position of the electrodes for each individual measurement, this way the
system will be able to memorize the electrical parameters measured for all the achieved
measurements that will be accepted by the operator.
Complete Measure
The complete measurement is made with a sequence of cycles, the results of these cycles
will be treated mathematically by Gea RM1 to provide the final result of the measurement
with the degree of accuracy required by the operator.
F
IGURE
27
-
EXECUTION OF A SINGLE MEASURE