background image

Instruction Sheet
for the PASCO
Model SF-9586

012-02815D

1/95

$1.00

Phone (916) 786-3800 • FAX (916) 786-8905 • email: [email protected]

10101 Foothills Blvd. • P.O. Box 619011 • Roseville, CA 95678-9011 USA

better

teach physics

ways to

© 1994 PASCO scientific
This instruction sheet edited by: Tom Reineking

To Use the Power Supply:

Plug the power cord into a well-grounded 107-128 VAC,
60 Hz or 198-242 VAC, 50 Hz outlet. (See label on
rear of unit).

Connect the appropriate output terminals of the power
supply to the circuit.  (Connecting wires are not pro-
vided with the supply.)

Turn the HIGH VOLTAGE ADJUST knob fully
counterclockwise.

Flip the ON/OFF switch on the front panel to ON.
The switch will light to show that the power is on.

Turn the HIGH VOLTAGE ADJUST knob to obtain
the desired output.  The meter will show the voltage
between the 6 KV output terminals.

CAUTION!—HIGH VOLTAGE

The power supply uses two 390 K

 resistors

which limit the output current to approximately
3mA.  Nevertheless:
* Always work with a partner.
* Wear rubber soled shoes.
* Work with one hand only.  Keep the other hand

in your pocket or behind your back.

KILOVOLT POWER SUPPLY

ON

OFF

SF-9586 

KILOVOLT POWER SUPPLY

HIGH VOLTAGE ADJUST

6.3 VAC

6 KV

3 KV

3 KV

CAUTION!

HIGH VOLTAGE

kV

Introduction

The PASCO scientific Model SF-9586  Kilovolt Power
Supply provides regulated DC power from 0 to 6 kilo-
volts.  The 6 kilovolt output is center tapped, providing
outputs of 0 to +/-3 kilovolts.  In addition, a pair of output
jacks provides 6.3 volts of AC power, which is commonly
used as filament power in electron tube experiments.

IMPORTANT:

The 6.3 VAC output is unregulated.  If you are

using it to heat the filament in  a vacuum tube, we
recommend that you use a current limiting resistor
(1-2 

, variable) in series.  Measure the voltage at

the filament and adjust the resistance to get the re-
quired voltage.

For an accurate meter reading, always use the

left and right output jacks (6 kV).  Use the middle
jack only when your application requires a center-
tapped voltage.  (The center tap will be the mid-
point of the output voltage only if equal currents
are drawn from the ±3 kV output jacks.)

Summary of Contents for SF-9586

Page 1: ...mA Nevertheless Always work with a partner Wear rubber soled shoes Work with one hand only Keep the other hand in your pocket or behind your back KILOVOLT POWER SUPPLY ON OFF SF 9586 KILOVOLT POWER SUPPLY HIGH VOLTAGE ADJUST 6 3 VAC 6 KV 3 KV 3 KV CAUTION HIGH VOLTAGE kV Introduction The PASCO scientific Model SF 9586 Kilovolt Power Supply provides regulated DC power from 0 to 6 kilo volts The 6 k...

Page 2: ...he warranty Shipping costs for returning the equipment after repair will be paid by PASCO scientific Other PASCO Power Supplies PASCO scientific offers a variety of rugged reliable power supplies From low voltage fixed output supplies to this 6KV DC supply all PASCO power supplies are easy to use ruggedly built and short circuit protected for long life in the student lab For information check our ...

Page 3: ... 777 M1 100ua KF86b P1A P1B 10K POT Q1 Q4 BC557B Q2 Q6 BD788 Q3 Q5 BD243C R4 R41 100R R7 R40 3K3 R8 12K4 1 R10 R31 22K R13 R15 47K R1 R36 2K43 1 R5 R42 22R 2W R6 R43 82R R9 R34 6K8 R11 R29 82K R12 R14 R47 R48 R49 10K R16 R19 R24 R28 10M R17 R26 8K2 R18 R32 1K R20 R33 75K 1 R21 R30 4K7 R23 2K2 R27 1M R2 274R 1 R22 R39 270R R3 R37 470R R25 R50 820R 5W R35 560R 1 R38 68K R44 R45 2K2 SIL R46 47K NTC R...

Page 4: ...4 Kilovolt Power Supply 012 02815D Layout Component Layout 1 of 2 ...

Page 5: ...5 012 02815D Kilovolt Power Supply Component Layout 2 of 2 ...

Page 6: ...6 Kilovolt Power Supply 012 02815D Schematic Schematic ...

Page 7: ...7 012 02815D Kilovolt Power Supply Schematic continued ...

Page 8: ...uld be helpful to prepare the following information If your problem is computer software related note Title and Revision Date of software Type of Computer Make Model Speed Type of external Cables Peripherals If your problem is with the PASCO apparatus note Title and Model number usually listed on the label Approximate age of apparatus A detailed description of the problem sequence of events In cas...

Reviews: