HPVA Series Operator’s Manual
Definitions of Fields in the System Windows
Mar 2011
2-29
Define Run Window
Equilibration Criteria
Specifies when the system has reached equilibrium after each step
in the adsorption/desorption procedure.
Enter the pressure to be reached and the number of minutes in
which it should be reached.
Enter the maximum equilibration time in minutes in the
Max
Equilib Time
field.
Control Method
Select the temperature control method to be used during the
experiment.
Click
Bath
if a recirculation bath or cryostat is to be used.
Click
External
if the furnace or a dewar is to be used.
Experiment
Temp
: Enter the temperature for the experiment in °C.
Hold
: Enter the number of minutes the system will wait once the
sample RTD reaches the experiment temperature. A hold time of
at least 60 minutes is recommended if a liquid nitrogen bath is
used.
Summary of Contents for HPVA Series
Page 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Page 4: ......
Page 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Page 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Page 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Page 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Page 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Page 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Page 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Page 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...