parkell TurboPIEZO Instructions For Use Manual Download Page 1

DEVICE DESCRIPTION

The Parkell TurboPIEZO™ is a piezoelectric 

ultrasonic tooth scaler with a high efficiency 

transducer that operates at a frequency of 

29KHz. It requires less cooling water and 

produces minimal noise. The TurboPIEZO has 

an expanded low-power range to improve 

patient comfort during debridement, and a 

power-boosting, user-controlled Turbo feature 

to increase scaling power when needed for 

heavy deposits.

INDICATIONS / INTENDED USE

For removal of calculus, plaque and oral debris 

from teeth during dental prophylaxis/scaling.

CONTRAINDICATIONS TO USE

If patient or operator is pregnant, or has any 

medical condition which might be affected by 

this device during treatment, consult a 

physician concerning advisability of performing 

dental scaling prior to use.

WARNINGS

•  This device should only be connected directly 

to a grounded electrical outlet.

•  The water supply to the scaler should always 

be turned off whenever the device is being 

connected, disconnected, or when not in use. 

• Do not run this device dry. Water must always 

flow during use to cool the handpiece and tip. 

•  The “ON/OFF” button on the front panel 

functions as the AC power disconnect device 

for the scaler. 

•  The transformer should be located above and 

away from any sources of water. As with all 

electrical devices, the unit should not be 

immersed in water or other liquids. Do not 

reach for the device if it has fallen into liquid 

until power is disconnected, and do not use 

the device after it has fallen into liquid. Return 

the device to Parkell for servicing. 

•  Modification of this device will void the 

warranty, and may violate safety codes, 

endangering the patient and/or the clinician.

•  This equipment is not suitable for use in the 

presence of a flammable anesthetic gas 

mixture (when used along with air or oxygen).

•  This equipment produces electromagnetic 

energy and may cause interference with other 

nearby electronic devices. Should this occur, 

changing the position or location of the 

device may be necessary.

 

WATER CONTROL

POWER CONTROL

When “PERIO” button is 

illuminated, this knob allows 

very precise low-power 

adjustments.

PERIO BUTTON

When depressed, the button 

illuminates indicating the 

scaler is in the low-power 

PERIO mode.

TURBO LIGHT

Indicates the device is operating in 

high-power Turbo mode (footswitch 

is fully depressed).

“ON” INDICATOR LIGHT

ON/OFF BUTTON

REMOVABLE  

HANDPIECE

PIEZO TIP

Tighten and remove with supplied 

Tip Wrench.

RX ONLY 

This manual, in whole or in part, should not be considered a substitute for formal training in Electrosurgery.  

Appropriate professional education is STRONGLY RECOMMENDED prior to using this device clinically.

INSTRUCTIONS FOR USE

A00436revD0320

PIEZOELECTRIC ULTRASONIC TOOTH SCALER  

h

 D690-110 (110VAC), 

h

 D690-230 (230VAC)

Summary of Contents for TurboPIEZO

Page 1: ...lectrical devices the unit should not be immersed in water or other liquids Do not reach for the device if it has fallen into liquid until power is disconnected and do not use the device after it has...

Page 2: ...Straight hD693 Perio hD695 Burnett Power Tip 1 Scaler Tip Wrench hD691 2 Inline water filters one attached to water line hD419 bag of replacement filters 1 Transformer with cables hD408 120V hD403 23...

Page 3: ...e OUT position CONTROLLING THE POWER OUTPUT The TurboPIEZO gives you several ways to adjust the scaling power 1 Use the Power Control Knob on the main power unit to set the initial scaling power for t...

Page 4: ...sink or cuspidor until desired mist is obtained For maximum patient comfort use a good flow of water as a coolant lubricant and to flush out debris Hold the handpiece in a comfortable pen grasp To re...

Page 5: ...the amount of water Increase water flow Excessive hand pressure applied Correct technique Heated water being used Use room temperature water only No water spray when foot pedal is depressed Water lin...

Page 6: ...1 249 1134 or email our Technical Support Service at techsupport parkell com WARRANTY For full Warranty and Terms of Use information please see www parkell com Parkell s Quality System is certified to...

Page 7: ......

Page 8: ...dental device was designed manufactured and is serviced in the United States of America by 300 Executive Drive Edgewood NY 11717 USA Toll Free 800 243 7446 Phone 631 249 1134 E mail info parkell com W...

Reviews: