Follow the steps below to prevent the nozzle from becoming
clogged:
• Reassemble the nebuliser.
• Switch the compressor on.
• Fill with fresh, soft water, boiled if necessary, or distilled wa-
ter, and nebulise it until no more aerosol comes out of the
nebuliser.
• Switch the compressor off again.
• Disassemble the nebuliser again.
• Pour out the collected water.
You can remove excess water more quickly by shaking all
parts.
NOTICE
The nozzle must be cleaned after every inhalation session,
otherwise the residues of the inhalation solution will quickly
crystallise in the nozzle and will
clog the nozzle
.
If the nozzle is clogged: [see: What should I do if ...,
page 45]
4.4 Disinfection
After cleaning, disinfect the dismantled nebuliser and accessor-
ies used (except for the connection tubing) (effective disinfec-
tion is only possible if the nebuliser has been cleaned).
Information when using masks with rubber bands:
The rubber band can only be cleaned as it might be dam-
aged by the high temperatures used during disinfection.
CAUTION
A damp environment may encourage the growth of bacteria.
Therefore, remove all parts from the pot or disinfector as
soon as disinfection has finished. Dry the parts. The
risk of
infection
is reduced when the parts are dried completely.
– 43 –
Summary of Contents for SOLE N Series
Page 1: ...SOLE N SOLE N Tracheo 2015 PARI GmbH Spezialisten f r effektive Inhalation 025D0070 A 10 15...
Page 3: ...de Gebrauchsanweisung 3 en Instructions for use 27...
Page 4: ......
Page 6: ...SONSTIGES 24 Entsorgung 24 Technische Daten 24 Zeichenerkl rung 25 Kontakt 25 4...
Page 28: ...26...
Page 30: ...MISCELLANEOUS 47 Disposal 47 Technical data 47 Explanation of symbols 48 Contact 48 28...
Page 51: ......