REPROCESSING IN HOME
ENVIRONMENTS
The product components must be cleaned thoroughly immedi-
ately after each application and disinfected at least once a
week.
The connection tubing cannot be cleaned or disinfected.
Dry the connection tubing after each use [see: Care of the
pressure gauge and connection tubing, page 24].
The maximum operating life of the connection tubing is 1 year.
4.1 Reprocessing cycles
PARI PEP S System
– Clean immediately after every use
– Disinfect once per week
Cleaning and disinfection can be carried
out together with the nebuliser.
Pressure gauge
Care as required [see: Care of the
pressure gauge and connection tubing,
page 24].
4.2 Processing limits
PARI PEP S System,
disinfection
300 processing cycles,
max. 1 year
– 19 –