Error codes
E 03
Internal hardware error (contact the Service Department).
E 04
Batteries are empty. This is displayed as soon as the batteries have insufficient capacity to carry
out measurements.
E 05
Measurement time has elapsed. This is displayed when measurement time (without inflation
time) exceeds 60 seconds.
E 06
This error code will be displayed when:
•
the actual cuff pressure exceeds the maximum allowable inflation pressure of 280 mmHg;
•
the device is not inflating beyond the set maximum pressure and is waiting for the next
measurement time.
E 07
The maximum inflation time of 60 seconds has elapsed. Error refers to a loose cuff or tube, or
another faulty sealing.
E 08
Insufficient oscillations have been detected:
At least 8 oscillations need to be detected for a correct measurement. The cuff should be
attached sufficiently tightly (you should be able to put one finger beneath the cuff, but not two).
PHYSIO-PORT will regulate the deflation rate for follow-up measurements accordingly.
E 10
Memory full. 200 blood pressure measurements have been carried out and the memory of the
device is full.
E 11
Movement during diastole recognition.
E 12
Diastole is outside of the measurement range.
E 20
Systole is outside of the measurement range.
(E12 and E20 are displayed when the calculated systole or diastole values are outside the
range in which the oscillations are recorded.)
E 21
Systole is below the measurement range.
E 22
Systole is above the measurement range.
E 23
Movement during systole recognition. Air deflation speed is too high, e.g. as a result of leakage.
E 24
Difference between systole and diastole is too small (10 mmHg or less)