Registering Inspection Results in Numerical
Calculation
(This is supported in Ver.1.30 or later.)
The detected Area of cells (for Connector (Binary Window)) or Gray Average (for Connector (Gray Window))
and Judgement can be registered into Numerical Calculation from the checker setting window. Although you
can register them also in Numerical Calculation setting window, information of many cells can be registered
efficiently with less operation from the checker setting window.
1.
Select "Num. Calcu.” in checker setting
window.
2.
Select “Yes” for “Num. Calcu.”
3.
Select the setting table and select a data type
to register.
Result:
Area or Gray average of each cell
Judgement:
Judgement of each cell
4.
Specify Starting Area, Number of Areas, and
No.
Starting Area:
Cell No. of the first cell to
register.
No. of Areas:
The number of cells to register
No.:
No. of the first Numerical
Calculation that is to be
registered.
In the setting table, always "0" are displayed.
Even after registering the data into Numerical
Calculation, the table shows “0” again.
The table cannot display the current registration
status of Numerical Calculation.
5.
Press the CANCEL key to go back to the
checker selecting window.
The data is registered at this timing.
When the destination Numerical Calculation No.
is already set an expression, a message is
displayed. Selecting "YES" overwrites the
expression.
The example shown above is the setting where the
results of cell No.0 to 4 are being registered into
Numerical Calculation No.3 to 7.
Summary of Contents for PV-500
Page 1: ......
Page 10: ......
Page 11: ...9 Names and Descriptions Chapter 1 INDEX Chapter 1 Names and Descriptions ...
Page 21: ...19 Installation and Wiring Chapter 2 INDEX Chapter 2 Installation and Wiring ...
Page 37: ...35 Basic Operation Chapter 3 INDEX Chapter 3 Basic Operation ...
Page 55: ...53 Setting Inspection Conditions Chapter 4 INDEX Chapter 4 Setting Inspection Conditions ...
Page 345: ...343 Chapter 7 Changing Configuration INDEX Chapter 7 Changing Configuration TOOL ...
Page 375: ...373 Operation and Data Output through I O Terminals Chapter 8 INDEX ...
Page 390: ......
Page 391: ...389 Control with Communication Port Chapter 9 INDEX Chapter 9 Control with Communication Port ...
Page 430: ......
Page 431: ...429 PVWIN Chapter 10 INDEX Chapter 10 PVWIN ...
Page 493: ......
Page 494: ......