VeriDri™ User’s Manual
7
Specifications
7.
Specifications
7.1
General Specifications
Moisture Ranges
• –110
o
to 20°C
• –110
o
to –50°C
• –80
o
to 20°C
• –80
o
to –30°C
• –30
o
to 20°C
• –60
o
to 40°C
• –150
o
to 70°F
• –150
o
to –40°F
• –40
o
to 70°F
• –100
o
to 0°F
• –50
o
to 50°F
• 0
o
to 100°F
• 0 to 10 PPMv
• 0 to 100 PPMv
• 0 to 1000 PPMv
• 0 to 10,000 PPMv
Note:
PPMv ranges are based on a constant pressure, which was provided at the time of the order.
Operating Temperature
–40° to 60°C (–40° to 140°F)
Storage Temperature
70°C (158
o
F) maximum
Note:
The probe should be stored with the plastic cover and desiccant packet threaded onto the probe. Store it in
a cool, dry environment.
Warm-up Time
Meets specified accuracy in 3 minutes
Calibrated Accuracy at 25°C (77°F)
• ± 2°C (3.6°F) from –65
o
to 10°C (–85
o
to 50
o
F)
• ± 3°C (5.4°F) from –80
o
to 66°C (–112
o
to 87
o
F)
Repeatability
• ± 0.5°C (0.9°F) from –65
o
to 10°C (–85
o
to 50
o
F)
• ± 1.0°C (1.8°F) from –80
o
to –66°C (–112
o
to –87
o
F)
7.2
Electronics
Power
•
Supply Voltage:
7 to 28 VDC (loop-powered, customer supplied)
•
Output:
4 to 20 mA
•
Output Resolution:
0.01 mA
Summary of Contents for VeriDri
Page 1: ...panametrics com 916 095 D Dec 2021 VeriDri Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 8: ...Contents vi VeriDri User s Manual no content intended for this page...
Page 18: ...Warranty 10 VeriDri User s Manual no content intended for this page...
Page 20: ...no content intended for this page...
Page 21: ......