Chapter 4. Error Codes
92
Sentinel™ User’s Manual
4.2 Error Descriptions and Actions
If an error message appears on the meter display screen during operation of the Sentinel, refer to the following list of
error descriptions for instructions on how to proceed.
4.2.1
Err 0: No Error
Problem:
No error condition currently exists.
Cause:
This message appears briefly to confirm that the response to another error message has corrected the
problem.
Action:
No action is required.
4.2.2
Err 1: Low Signal
Problem:
Poor ultrasonic signal strength or the signal exceeds the limits entered in the New Meter Browser setup
(see Chapter 2,
Initial Setup
).
Cause:
Poor signal strength may be caused by a defective cable, a flowcell problem, a defective transducer or a
problem in the electronics console. A signal that exceeds the programmed limits is probably caused by
the entry of an improper value in the
New Meter Browser
information setup (see Chapter 2,
Initial Setup
).
Action:
Using the procedures in Chapter 5,
Diagnostics
, check the components listed above. Also, check the
value entered into the New Meter Browser information setup (see Chapter 2,
Initial Setup
).
4.2.3
Err 2: Soundspeed Error
Problem:
The soundspeed exceeds the limits programmed in the New Meter Browser information setup (see
Chapter 2,
Initial Setup
).
Cause:
The error may be caused by incorrect programming, poor flow conditions or poor transducer orientation.
Action:
Compare the measured soundspeed to tabulated nominal values for the fluid being used and correct
any programming errors. Refer to Chapter 5,
Diagnostics
, to correct any flowcell and/or transducer
problems.
Summary of Contents for Sentinel
Page 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents x Sentinel User s Manual...
Page 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Page 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Page 90: ...Chapter 3 Operations Sentinel User s Manual...
Page 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Page 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Page 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Page 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Page 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Page 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Page 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Page 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Page 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Page 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Page 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Page 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Page 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Page 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Page 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Page 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Page 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Page 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Page 158: ...Index 158 Sentinel User s Manual...
Page 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Page 162: ......
Page 164: ......
Page 165: ......