Chapter 1. Installation
6
Sentinel™ User’s Manual
1.5.5 Maximum and Minimum Flow
Maximum and minimum flow rates through the
Sentinel Flow Measurement System
are based on the pipe diameter
and the process fluid pressure. The information in the following tables is approximate, and is based on representative
natural gas components at a process temperature of 70°F (21°C). See
Table 3
and
Table 4
below for English units and
Table 5 on page 7
and
Table 6 on page 7
for metric units.
psig
6”
8”
10”
12”
14”
16”
18”
20”
24”
100
16.2
28.0
44.2
47.3
57.2
74.7
94.5
117.5
169.9
200
30.7
53.2
83.9
89.8
108.6
141.8
179.5
223.1
322.6
400
61.1
105.9
166.8
178.6
215.9
282.0
357.0
443.6
641.6
600
93.3
161.5
254.6
272.6
329.5
430.4
544.8
677.0
979.1
800
127.3
220.4
347.4
371.9
449.4
587.1
743.2
923.4
1335.6
1000
163.0
282.2
444.9
476.3
575.6
751.9
951.8
1182.7
1710.5
1200
200.3
346.8
546.7
585.3
707.3
924.0
1169.6
1453.3
2102.0
1400
238.8
413.6
651.9
697.9
843.5
1101.9
1394.7
1733.1
2506.6
1480
254.5
440.7
694.7
743.7
898.8
1174.1
1486.2
1846.8
2671.0
Maximum flow rates are based on 118 ft/sec flow velocity for 6” through 10” diameter pipes,
and on 89 ft/sec for 12” through 24” diameter pipes.
psig
6”
8”
10”
12”
14”
16”
18”
20”
24”
100
0.3
0.6
0.9
0.8
1.0
1.2
1.6
2.0
2.8
200
0.6
1.1
1.7
1.5
1.8
2.4
3.0
3.7
5.4
400
1.2
2.1
3.3
3.0
3.6
4.7
5.9
7.4
10.7
600
1.9
3.2
5.1
4.5
5.5
7.2
9.1
11.3
16.3
800
2.5
4.4
6.9
6.2
7.5
9.8
12.4
15.4
22.2
1000
3.3
5.6
8.9
7.9
9.6
12.5
15.8
19.7
28.4
1200
4.0
6.9
10.9
9.7
11.8
15.4
19.4
24.2
35.0
1400
4.8
8.3
13.0
11.6
14.0
18.3
23.2
28.8
41.7
1480
5.1
8.8
13.9
12.4
14.9
19.5
24.7
30.7
44.4
Minimum flow rates are based on 2.36 ft/sec flow velocity for 6” through 10” diameter pipes,
and on 1.48 ft/sec for 12” through 24” diameter pipes.
Summary of Contents for Sentinel
Page 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents x Sentinel User s Manual...
Page 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Page 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Page 90: ...Chapter 3 Operations Sentinel User s Manual...
Page 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Page 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Page 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Page 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Page 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Page 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Page 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Page 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Page 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Page 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Page 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Page 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Page 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Page 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Page 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Page 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Page 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Page 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Page 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Page 158: ...Index 158 Sentinel User s Manual...
Page 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Page 162: ......
Page 164: ......
Page 165: ......