Sentinel LNG User’s Manual
41
Chapter 3. General Programming
3.4.2 Locking the Sentinel LNG
The security level can be accessed in two ways.
From the display screen:
1.
Press the [u] key three times, until the lock in the upper right corner is highlighted.
2.
Press
[ENT]
, and proceed to step 4 below.
From the User Program:
1.
Press
[ESC]
. The
Sentinel LNG
enters the User Program.
2.
Press the [u] key until USER is bracketed.
3.
Press the [q] key until Set Security is highlighted. Press
[ENT]
.
4.
The screen shows up to ten options (depending on the current security level):
• Full Lock – This will be the default level during meter power-up. Any User who wishes to access the menu will
have to press the keys in the following combination (
Escape
-
Enter
-
Escape
) to unlock the meter with a “Valid
Passcode”. The entered “Passcode” is validated and the user is allowed to access the menu based on
privileges. The meter will go back to Full-Lock mode if no keypad activity is noted for a “Determined Time”.
• Display Locked – This level can be achieved by entering a “Valid Passcode” assigned for this role. In this
Mode/Level, the User has privilege only to view data. The User will be allowed to select the diaplay parameters.
Any other changeable parameter is denied access.
• User-1 / User-2 / User-3 / User-4 – The User has to enter a “Valid Passcode” to have access to the menu
attached to this role. The User with this role will have rights of Display Locked mode and some predetermined
“write privilege” with which he may modify parameters (attached to this role).
• User Admin – This User will have “Administrator privilege” in addition to privileges of Lower level. This user is
given the rights to change the “Passcode” for all user-levels. (This does not include Full Lock and Display
Locked). The Administrator has the privilege to modify the timeout value, which is used to determine the time
taken to switch back to Full Lock mode (if no keypad activity is noted). The “Administrator privilege” does not
include access to Service or Factory Menu. The Administrator has no rights or control on
Service/Factory/Engineering Levels or their “Passcodes”.
• Field Service – This User has access to the “Service Menu” in addition to “Administrator privilege”. It is not
accessible by User Admin.
• Factory – This User has access to “Factory privilege” in addition to “Service privileges”. It is not accessible by
User Admin.
• Engineering – This User is the only user with Full access to the meter. Only this user has rights to change the
“Passcodes” for Factory and Service. It is not accessible by User Admin.
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......