TransPort™ PT878GC User’s Manual
9.1.4.4
Calibrating the Thickness Gauge Transducer
To zero the transducer offsets, go to the Zero tab. Press the [
] or [
] arrow key until you reach the Zero tab and
press
[ENTER]
. The screen appears similar to Figure 137 below.
Figure 137: The Zero Window
1.
Press the [
] arrow key to enter the window.
2.
The first prompt asks you to choose between single and dual-point calibration. Use the [
] and [
] arrow keys
to move to the appropriate radio button and press
[ENTER]
.
Note:
Panametrics Sensing recommends dual calibration. Dual-point calibration is more reliable, but it requires
two known thicknesses of the calibration material.Use single-point calibration only if you know the sound
speed in the calibration block material to a high degree of accuracy. See Appendix C for more information.
3.
The next prompt asks you to enter the length of Block 1. (If you are using a Panametrics Sensing-supplied test
block, the length is printed on the block.) Press
[ENTER]
to open the text box.
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the length.
c.
Press
[ENTER]
to confirm the entry.
4.
Press
[ENTER]
(Calibrate button) to prepare for calibration. Then apply couplant B to the block surface, and hold
the transducer against the block.
5.
When you have the transducer firmly pressed against the block, press
[ENTER]
(Ready button) again to perform
the calibration.
6.
Hold the transducer steady until the Set button appears. Press
[ENTER]
to store the calibration.
IMPORTANT:
Be sure to hold the transducer steady until the Panametrics Sensing wait cursor disappears.
7.
If you have chosen dual calibration, repeat Steps 3 through 6 above for Block 2.
8.
The PT878GC asks for confirmation of the calculated and current values.
• Press the [
] arrow key to move to the Retry box and press
[ENTER]
to recalculate the value, or
• Press
[ENTER]
on “Set” to commit the calculated value.
Once you have completed calibration, the transducer should remain calibrated until the meter memory is cleared or
a dramatic shift in ambient conditions occurs.
To leave the Zero window:
Display
Thickness Gauge Measure
Cancel
OK
Graph
Velocity
Zero
Material
Block 1 Length
Calculated
Current
Transducer Delay,
0
0
Single
Dual Calibration
0 mm Calibrate
µs
Summary of Contents for PT878GC
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii TransPort PT878GC User s Manual...
Page 16: ...Chapter 2 6 TransPort PT878GC User s Manual Figure 3 Connection Locations...
Page 50: ...Chapter 3 40 TransPort PT878GC User s Manual...
Page 98: ...Chapter 5 88 TransPort PT878GC User s Manual...
Page 148: ...Chapter 8 138 TransPort PT878GC User s Manual...
Page 152: ...Chapter 9 142 TransPort PT878GC User s Manual Figure 131 Printout of a Typical Drive Report...
Page 190: ...Chapter 10 180 TransPort PT878GC User s Manual...
Page 196: ...Chapter 11 186 TransPort PT878GC User s Manual...
Page 198: ...Appendix A 188 TransPort PT878GC User s Manual...
Page 199: ...Appendix A TransPort PT878GC User s Manual 189 Figure 163 The Site Menu...
Page 201: ...Appendix A TransPort PT878GC User s Manual 191 Figure 165 The Program Menu Pipe Option...
Page 204: ...Appendix A TransPort PT878GC User s Manual 194 Figure 168 The Meter Menu...
Page 205: ...Appendix A TransPort PT878GC User s Manual 195 Figure 1 Logging Mneu...
Page 212: ...Appendix D 202 TransPort PT878GC User s Manual...
Page 218: ...Index 208 TransPort PT878GC User s Manual...
Page 220: ...2 Doc Tittle Manual Type...
Page 222: ...Warranty 4 TransPort PT878GC User s Manual no content intended for this page...
Page 223: ......