A.9.30 Command 171 (0xAB): Read Output Trim Zero
This command is to trim the zero or low endpoint value to a value of 4 mA.
For example, if the zero value is 4.1 mA, then the correction entered should be -0.1 mA to achieve a result of 4 mA.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
A.9.31 Command 171 (0xAC): Read Output Trim Span
This command is to trim the zero or low endpoint value to a value of 20 mA.
For example, if the zero value is 20.1 mA, then the correction entered should be -0.1 mA to achieve a result of 20 mA.
Request Data Bytes
6
Error
Device-Specific Command Error
7
Error
In Write Protect Mode
8-15
Undefined
16
Error
Access Restricted
8-127
Undefined
Byte
Format
Description
0
Byte
Format
Description
0
Unsigned-8
Enumeration. 0 = Fixed 4 mA trim, 1 = Live mA
Read for live analog output or fixed 4 mA for trim purposes
1~4
Float
mA correction Value
Code
Class
Description
0
Success
No Command-Specific Errors
1-4
Undefined
5
Error
Too Few Data Bytes Received
6
Error
Device-Specific Command Error
7
Error
In Write Protect Mode
8-15
Undefined
16
Error
Access Restricted
8-127
Undefined
Byte
Format
Description
0
Code
Class
Description
Summary of Contents for HydroProII
Page 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Preface x HygroProII User s Manual no content intended for this page...
Page 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Page 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Page 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Page 103: ......